OPERATOR
MAINTENANCE
TROUBLESHOOTING
Sweep Generator Quits During Operation
(
OPERATE
light remains on)
Trouble Description: The sweep generator operates for some time,
then shuts down (
OPERATE
light remains on). After a short period, the
sweep generator resumes normal operation. This is an indication that
the 681XXA has reached an excessive operating temperature.
Step 1.
Check that the fan is still operating during the time that the
instrument is shut down.
If the fan is still operating, clean the air filter (see page
6-14).
If the fan is not operating, call a service technician.
One or More Error Messages
are Displayed During Start Up
Step 1.
Perform a self-test of the sweep generator by pressing the
System Menu soft-key
Selftest
.
If error message (s) does not repeat, resume normal opera-
tion.
If error message(s) remains, call a service technician.
LOCK ERROR
is Displayed
Step 1.
Perform a self-test of the sweep generator by pressing the
System Menu soft-key
Selftest
.
If self-test does not result in an error message(s), resume
normal operation.
If an error message(s) is displayed, call a service techni-
cian.
Table 6-4.
Troubleshooting (2 of 3)
681XXA OM
6-11