Selecting an Alternate Sweep Range
To go to the Alternate Range menu (shown below)
from the Alternate Sweep menu, press
Alt Range
.
Select the alternate sweep range (
Full, F1-F2
,
F3-F4
,
F5-dF
, or
F6-dF
). The menu then displays the cur-
rent frequency parameters for the selected sweep
range. If you wish to change a frequency parameter,
use the menu edit soft-key to open the parameter,
then edit it.
When you are done selecting the alternate sweep
range, press
Previous Menu
to return to the Alter-
nate Sweep menu.
Selecting an Alternate Sweep Power Level
To go to the Alternate Level menu (shown below)
from the Alternate Sweep menu, press
Alt Level
.
Select the power level for the alternate sweep range
(
L0, L1
,
L2
,
L3
, or
L4
). The menu then displays the
current level parameter for the selected power level.
If you wish to change the level, use the menu edit
soft-key to open the parameter, then edit it.
A menu edit soft-key is also provided to let you
change the power level of the main sweep.
When you are done selecting the power level for the
alternate sweep range and editing the power level
of the main sweep, press
Previous Menu
to return
to the Alternate Sweep menu.
LOCAL (FRONT
SWEEP FREQUENCY
PANEL) OPERATION
OPERATION
CAUTION
Performing alternate sweeps using
power levels that cross step attenu-
ator switch points can cause exces-
sive wear on the switches and re-
duce the life expectancy of the step
attenuator.
681XXA OM