background image

             Installation  and operating  instructions  WILO Mather and Platt 

MISO / PISO Pumps 

20  

 

 

 
8.3.2    Pre-starting  checks  (Refer  Check  List  in 
Annexture-2) 

 

Check that the inlet isolating valve is open, and that the 
delivery valve is closed. 

 

Check that there is no blockage in the strainer at the end 
of the suction line. 

 

Check for free rotation of the unit when coupled. 

 

Check  that  suction  and  delivery  pressure  gauges  are 
connected. Test and make available any alarm, signals, 
interlock  systems  and  any  of  the  protective  devices 
incorporated in the auxiliary and main pumping control 
system. 

 

Ensure that all electrical checks on motor, relay setting 
in panel etc. have been carried out in accordance with 
the instructions of motor manufacturer. 

 

Ensure that stuffing box sealing water seal connection is 
provided as shown in GA Drawing. 

 
          8.3.3 Normal starting and running checks 

 

 

When all the foregoing pre-start checks are satisfactory, 
start  the  pump  and  check  the  direction  of  rotation 
(indicated  by  a  direction  arrow  on  the  pump  casing) 
otherwise stop the pump immediately for correction of 
direction  of  rotation.  Then  run  the  pump  at  its  rated 
speed. 

 

Check the ammeter reading to ensure that the motor is 
not being overloaded. 

 

If  applicable,  ensure  that  the  stuffing  box  is  not 
overheating  and  that  there  is  slight  leakage  from  the 
gland (about 1 -2 drop per second). There may be at first 
a tendency for the stuffing boxes to run warm because 
of the high viscosity lubricant in the packing. During the 
first few minutes of running with new packing, a small 
quantity of very viscous fluid will be extruded, but the 
flow should reduce when the packing has settled down. 

 

Check  the  mechanical  seal  for  leak.  In  the  start  phase 
(and  also  after  downtimes)  slight  leakage  can  be 
expected.  Visual  leakage  checks  are  however  required 
from time to time. Distinctly visible leakage will require 
an exchange of the seal. WILO Mather and Platt offers a 
repair set containing all parts required for an exchange. 

 

Check that the bearing is not overheating. Bearings will 
norm

ally  run  at  a  temperature  of  30˚C

-

35˚C  above 

ambient temperature. The ideal running temperature of 

bearings  is  40˚C  to  60˚C  for  ball  bearings  and  40˚C  to 
55˚C for bush bearings. The temperature should never 
exceed  82˚C  for  ball  bearings  and  75˚C  for  bush 

bearings. If the bearings are overheating its cause should 
be investigated immediately. 

 

 If  the  foregoing  checks  are  satisfactory,  open  the 
delivery valve slowly and bring the pump gradually up to 
its  rated  parameters  indicated  in  the  data  sheet/name 
plate  and  based  on  pressure  gauge  and  ammeter 
readings. Unless the pump is fitted with a special leak-
off device, it should not be run for a long period against 
a closed delivery valve. Check that the driving unit is not 
being  overloaded  during  valve  opening.  Overloading 
may  occur  if  the  pump  is  discharging  into  an  empty 
system. If the pumping unit fails to generate at least its 
rated delivery pressure it must be stopped immediately, 
the cause ascertained, 

 

Check vibration of pump set and ensure that vibration 
level is within limits specified. Check that noise level is 
within stipulated limits. 

 
 

 

The pumps may be run for 8 hours trial operation and all the 
parameters  like  delivery  pressure,  current,  bearing 
temperature, etc. Be recorded periodically. 

 

Make  the  following  checks  at  regular  intervals.  It  is 
recommended that they be made at every change of shift. 

 

Check the suction and discharge pressure gauge for normal 
operating pressure, if there is significant drop in the suction 
or discharge pressure the pump may have lost its supply. In 
the event of this fault occurring, the pump must be stopped 
immediately, and the cause of liquid loss eliminated. 

 

Check  the  mechanical  seal  or  stuffing  box  assembly  for 
overheating. 

 
                   8.3.4 Sealing system 

            

             Gland packing 

 
            CAUTION! Risk of damaging the pump! 
If  the  gland  plate  is  too  tight,  the  packing  stuff  will  be 
immediately damage. 

At  the  beginning  of  the  operation,  the  leak  at  the  gland 
packing should be important. It should reduce progressively 
after several hours by a balanced and reasonable tightening 
the  gland  plate.  The  gland  packing  must  operate  without 
excessive  temperature.  The  correct  setting  of  the  gland 
packing  let  a  permanent  leak  around  1  or  2  drops  per 
seconds. 
If  this  leak  is  too  much  and  cannot  be  adjusted  with  the 

gland  plate,  the  packing  stuffing’s  are  worn  and  must  be 

replaced. 
 

          Mechanical seal 
 

            CAUTION! Risk of damaging the pump! 
A mechanical seal must never operate without fluid and 
lubrication even for a short period of time. 

ensure that the pump is completely full of water and vented 
before starting the pump. Small leakages can occur during 
the period of running-in, they should disappear after several 

hours of operation. If the leakages don’t stop, shut down the 

pump  disassemble  the  mechanical  seal  and  control  their 
condition. 

 

                   8.3.5 Normal shutdown 
 

            WARNING! Risk of Burns! 
If the fluid temperature and system pressure is high, close 
the  isolation  valves  upstream  and  downstream  of  the 
pump. Initially let the pump cool. 

 

Close  the  delivery  valve  to  reduce  the  load  on  the  driving 
unit. 

 

Stop the driver of the pump. 

 

When  the  pump  has  come  to  rest,  close  the  suction-
isolating valve. 

 

Isolate any ancillary supplies. 

 

                   8.3.6 Emergency Shutdown 

In the event of any malfunction of the equipment, switch off 
the pump set. When the pump has come to rest, close the 
suction  &  discharge  valves,  isolate  the  driving  unit  power 
supply & rectify the fault. 

Summary of Contents for MISO 100-200

Page 1: ...WMPPPL_IOM_MISO_PISO_RV 003_20201002 WILO Mather and Platt MISO PISO Back Pull Out Pumps en Installation and Operating Instructions...

Page 2: ...e product at site The authorities that install and maintain the pump shall be responsible for hassle free installation operation or maintenance of the product This document is prepared with at most ca...

Page 3: ...Pumps 3 Fig 1 Pump Handling Bare Pump and Pump with Motor Engine Page No 9 Point No 3 1 Fig 2 Engine Driven Pump Page No 11 Point No 6 1 2 Fig 3 Pump Foundation Page No 15 Point No 7 2 1 Equal Clearan...

Page 4: ...LO Mather and Platt MISO PISO Pumps 4 Fig 4 Leveling of Base plate Page No 15 Point No 7 2 2 Fig 5 Coupling assembly Page No Point No 7 2 3 Fig 5 Coupling Alignment Page No 16 Point No 7 2 3 Fig 6 For...

Page 5: ...structions WILO Mather and Platt MISO PISO Pumps 5 Fig 7 Suction line Good and Bad layout Page No 17 Point No 7 2 5 Suction line God Page No 17 Point No 7 2 5 Page No 17 Point No 7 2 5 Fig 8 Suction l...

Page 6: ...Installation and operating instructions WILO Mather and Platt MISO PISO Pumps 6 Fig 11 Stuffing box packing Page No 18 Point No 7 2 7...

Page 7: ...roductinformation 10 5 1 Data plate 10 5 2 Type key 11 5 3 General description 11 5 4 Scope of delivery 11 5 5 Accessories 11 6 Description andfunction 11 6 1 Description of the product 11 7 Installat...

Page 8: ...e if this information is disregarded CAUTION There is a risk of damaging the pump installation Caution implies that damage to the product is likely if the information is disregarded NOTE Useful inform...

Page 9: ...Disposal of any unwanted scrap material should be disposed in appropriate way so as not to cause any harm to the environment No hazardous material is used in WILO Mather and Platt MISO PISO pumps NOT...

Page 10: ...ed at regular intervals This prevents the bearing from locking and the film of lubricant on the mechanical shaft seal is renewed For mechanical seal we recommend relative air humidity below 65 tempera...

Page 11: ...ate 5 5 Accessories Companion Flange Foundation bolts Shims Pressure Gauges Relief Valves 6 Description and function 6 1 Description of the product MISO PISO pumps are horizontal single stage volute c...

Page 12: ...y be fitted in the stuffing box situated at each end of the casing Gland Pack For correct operation of the pack gland as light leakage 40 60 drops per minute is essential If the leakage is reduced too...

Page 13: ...of MISO pump consist of following parts For glandpack version pumps No Part description No Part description 1700 Impeller nut 2310 Gland 1501 Impeller 2070 Thrust collar 2111 Spacer sleeve 2600 Bearin...

Page 14: ...coupling must comply with the standard EN349 The alignment of the pumps and the motor must be checked after the installation of the pump set on its foundation and when the piping is connected In addit...

Page 15: ...olts For installations where a low level of noise is expected built the foundation in a pit lined with appropriate insulation material in order to avoid vibration transmission to the ground CAUTION Ri...

Page 16: ...un freely and that the shaft is true to 0 1mm or better Check that no damage can be caused when the shaft of the driven unit is turned Coupling should be loosely coupled and the halves must be free to...

Page 17: ...Unequal nominal widths of the suction branch and suction line must be compensated by eccentric transition pieces refer figure 8 Pg No 5 It is recommended that a strainer is installed in front of the...

Page 18: ...ion manual at the time of installation and connection Motors or electrical control panels are operated with alternating or industrial high voltage current The electrical connection is established via...

Page 19: ...e closed after priming and before the pumping set is started 8 2 2 Pumps operating on negative suction head There are two methods of priming pumps that draw their liquid from an elevation lower than t...

Page 20: ...with a special leak off device it should not be run for a long period against a closed delivery valve Check that the driving unit is not being overloaded during valve opening Overloading may occur if...

Page 21: ...per cooling and lubrication if applicable For mechanical seal check that there is no visible leakage For any abnormality observed from the visual manual inspection and through maintenance inspection l...

Page 22: ...equire renewal can be made NOTE The figures given in the table Technical Data Pg No 12 above are only valid if the wear rings and the impeller are made with in the same materials of low galling tenden...

Page 23: ...1185 on bench Insert the packing 2410 logging ring 2300 set into the stuffing box Fit the gland 2310 into stuffing box Hold the gland with nuts finger tight Fasten the back cover assembly to frame ada...

Page 24: ...nt and back side of the impeller Assemble the pin and bush coupling 4630 onto the shaft 1800 Fit all studs and nuts of the coupling frame adaptor and motor stool Assemble the suitable motor over the m...

Page 25: ...se of damages exchange mechanical seal 11 Pump or pipe work not completely filled Vent and fill the pump as well as the suction line or in flow line 12 Pump is running against closed valve Open the sh...

Page 26: ...ISO Grade 46 oil Manufacturer Bush Type Bearings Grease Lubricated Bearings Indian Oil Corporation Servo System 46 Servogem 2 3 Hindustan Petroleum Enclo 46 Lithion 2 3 Bharat Petroleum Hydral 46 Mult...

Page 27: ...Neck ring 14 2600 Bearing 3 1185 Back cover 15 2752 Bearing end cover NDE 4 1192 Support foot 16 2753 Bearing end cover DE 5 1501 Impeller 17 2760 Bearing housing 6 1700 Impeller nut 18 2902 Internal...

Page 28: ...2 1181 Neck ring 15 2752 Bearing end cover NDE 3 1185 Back cover 16 2753 Bearing end cover DE 4 1192 Support foot 17 2760 Bearing housing 5 1501 Impeller 18 2902 Internal circlip 6 1700 Impeller nut...

Page 29: ...End cover DE 4 1192 Support foot 18 2760 Bearing Housing 5 1501 Impeller 19 2902 Internal Circlip 6 1700 Impeller nut 20 2902 1 External Circlip 7 1800 Shaft 21 3200 Key Impeller 8 1950 Oil Thrower 22...

Page 30: ...d cover DE 3 1185 Back cover 16 2760 Bearing housing 4 1192 Support foot 17 2902 Internal circlip 5 1501 Impeller 18 2902 External circlip 6 1700 Impeller nut 19 3200 Key impeller 7 1800 Shaft 20 3250...

Page 31: ...2 Support Foot 19 2902 1 External Circlip 5 1501 Impeller 20 3200 Key Impeller 6 1700 Impeller Nut 21 3250 Key Coupling 7 1800 Shaft 22 4031 Lip Seal 8 1950 Oil Thrower 23 4250 O ring for Sleeve 9 207...

Page 32: ...Adaptor 19 2760 Bearing Housing 5 1500 Impeller 20 3011 Lock Nut 6 1650 Impeller Nose Cap 21 3020 Lock Washer 7 1800 Pump Shaft 22 3200 Key for Impeller 8 1822 Adjusting Ring 23 3250 Key for Coupling...

Page 33: ...ushing sealing liquid for stuffing box 9 Availability of sufficient cooling liquid for bearings as specified 10 Free rotation of pump and drive shafts 11 Lubrication of bearings 12 Checking of insulat...

Page 34: ...fied 12 Free rotation of pump and drive shafts 13 Batteries are fully charged 14 Battery cables and lead are available 15 Exhaust silencer and all required exhaust piping is completed 16 Fuel tank is...

Page 35: ...Voltage Current Details of System Piping Details Application Suction Pipe Size Liquid Delivery Pipe Size pH Value Valves Suction Flooded Lift Expansion Joints Pump Operating Parameters Motor Operating...

Page 36: ...System Details of System Piping Details Application Suction Pipe Size Liquid Delivery Pipe Size pH Value Valves Suction Flooded Lift Expansion Joints Pump Operating Parameters Engine Operating Parame...

Page 37: ...lation and operating instructions WILO Mather and Platt MISO PISO Pumps 37 Pump set Levelling Record Required accuracy of pump set levelling is 0 05 mm Result Customer Representative WILO M P Represen...

Page 38: ...Pump and Motor Engine Turbine Alignment Record A Before Connecting Suction and Delivery Piping Gap Between Coupling Halves mm Actual Gap mm B After Connecting Suction and Delivery Piping Gap Between...

Page 39: ...latt MISO PISO Pumps 39 Pump set Noise Level Record Noise level measurement at Pump House Floor Noise Level in dBA A B C D Background Noise dBA Instrument Details Distance 1 Mtr from the Pump Make Mod...

Page 40: ...tails Date Pump Sr No Motor Sr No Pump Type Make Project Vibration Measurement Pk to Pk Sr No Position V H A Displacement Velocity mm sec Displacement Velocity mm sec Displacement Velocity mm sec 1 Pu...

Page 41: ...g Temperature Measurement Pump Set Details Date Pump Sr No Motor Sr No Pump Type Make Project Bearing Temperature Sr No Position Temperature Ambient Temperature Remarks 1 Pump DE 2 Pump NDE 3 Motor DE...

Page 42: ...Installation and operating instructions WILO Mather and Platt MISO PISO Pumps 42 NOTES...

Page 43: ...Installation and operating instructions WILO Mather and Platt MISO PISO Pumps 43 NOTES...

Page 44: ...________________________________________________________________________________________________ WILO Mather and Platt Pumps Pvt Ltd Mumbai Pune Road Chinchwad Pune 411 019 Maharashtra India Tel 91 20...

Reviews: