6. Faults
6. Faults
CAUTION!
Physical injuries, damage to equipment
If faults cannot be eliminated by means of the measures listed, the tensi-
on/compression force transducer must be taken out of operation immedi-
ately.
▶
Contact the manufacturer.
▶
If a return is needed, please follow the instructions given in chapter 9.2
“Returns”.
For contact details, please see chapter 1 “General information” or the back
page of the operating instructions.
Faults
Causes
Measures
No display
Leaking of the hydraulic force
measuring unit or pressure
gauge
Visual check whether the
transmission medium has
leaked (glycerine/water
mixture)
Consult the manufacturer
Deviation of the zero point
display
Overload, load offset, defecti-
ve force introduction
Consult the manufacturer
Output signal constant
when changing force
Mechanical overload, defecti-
ve force introduction
Consult the manufacturer
Signal span varies
EMC interference sources in
the environment; for example,
frequency converter
Shield instrument; cable
shield; remove source of
interference
Signal span drops/too small Mechanical overload, leaking
of the hydraulic force measu-
ring unit
Consult the manufacturer
Pointer jerks with force
increase
Stick-Slip effect (internal
pressure too low)
Consult the manufacturer
6.1 Faults of the hydraulic force measuring unit with pressure gauge / digital
pressure gauge