9037926
07/2015
61
WIKA Operating Instructions
CTD9300
Rev. 1
GB
Temperature Dry Well Calibrators
CTD9300
5. Technical Data
5.2
CTD9300
from 40 to 650 °C
Minimum temperature
Maximum temperature
Error of measurement, 100 °C/
within the range of measurement
Error of measurement 100 °C
Stability up to 100 °C
Stability at 650 °C
Gradients axially, 4 cm, 650 °C
Gradients axially, 4 cm, 100 °C
Heating time, 20 °C to 650 °C
Cooling time, 650 °C to 100 °C
Immersion depth
Sleeve diameter
Resolution in K
Footprint W x D
Footprint
Dimensions, W x D x H
Power consumption
Power supply
Fuses
Measurement value logging
Mass (without integ.
measuring instrument)
Ambient temperature
Standard ambient conditions
Overvoltages
Degree of pollution
Storage temperature
Humidity
Protection class
CE conformity
°C
°C
K
%
K
K
K
K
min
min
mm
mm
K
mm
cm²
mm
W
VAC
kg
°C
°C
% rel.h.
40
650
0.1
0.1
0.03
< 0.1
< 0.4
< 0.06
30
100
150
28
0.01
160 x 320
512
160 x 320 x 420
1,000
230 VAC, +10%/-15%, 50/60 Hz
115 VAC, +10%/-15%, 50/60 Hz, opt.
6.3 A, slow blowing (T6.3 A 250V) at
230 VAC; 10 A, slow blowing
(T10 A 250V) at 115 VAC
optional
10
0...40
Use permissible indoors only
permissible altitude up to 2,000 m
Overvoltage category II
Transient overvoltageas it normally
occur in the supply network
Degree of pollution 2
-20 ... 40
0 ... 90
IP 20
EN61326-1, EN61010-1
Summary of Contents for CTD9300 Series