9037926
07/2015
Temperature Dry Well Calibrators
CTD9300
3. Moduses and Functions
GB
40
WIKA Operating Instructions
CTD9300
Rev. 1
This menu is used to set the control parameters.
3.1.3.3 Controler
Fig. 25: Control parameter set
The parameter set in the basic configuration is as follows (device-
dependent):
Control parameter
CTD9300
-650
P0
1.00
P1
1.00000E-03
P2
1.00
P3
0.00
P4
1.00000E-01
CTD9300
-165
1.00
1.00000E-02
1.00
0.00
1.00000E-01
Where:
Control parameter
Designation
P0
internal amplification
P1
adjustment of the disturbance variable
P2
time constant
P3
test parameter
P4
safety limit of the dynamic
Warning!
Ex works, the control parameters are set for an optimum
control behaviour.
The user should not edit these settings.