9037926
07/2015
GB
Temperature Dry Well Calibrators
CTD9300
3. Moduses and Functions
49
WIKA Operating Instructions
CTD9300
Rev. 1
3.3.2.3 Calibration Menu “Pt-Probe”
In this menu, the output signal
of a Pt resistance thermometer
can be displayed in addition to
the nominal and actual
temperatures of the calibrator.
Fig. 36: Calibration menu
“Pt-probe”
The following information is provided:
Designation
Probe
t-Probe
Error
Signal
Information
Type of probe
Temperature of the
probe
Error of the probe
in K
Electric output signal
of the probe
Note
Pt resistance thermometer:
Pt 100-3, i.e.
Pt 100 in 3-wire circuits
in this example: 102.99 °C
Deviation from the int. (ext.)
reference thermometer
In the case of the Pt resistance
thermometer, the resistance in
Ohm