9037926
07/2015
Temperature Dry Well Calibrators
CTD9300
3. Moduses and Functions
GB
38
WIKA Operating Instructions
CTD9300
Rev. 1
This menu comprises the parameters for linearising the internal
reference thermometer.
The internal reference
thermometer is linearised in two
stages. First of all, using a basic
linearisation that is always
performed and thereafter using
an individual fine linearisation.
In this case, "t'" is the
temperature following the basic
linearisation.
t
display = P0 + P1 * t' + P2 * t' ² + P3 * t' ³ + P4 * t' 4
Linearisation
CTD9300
-650
CTD9300
-165
parameters
P0
0
0
P1
1
1
P2
0
0
P3
0
0
P4
0
0
Using these parameters, only the general basic linearisation is
enabled and fine linearisation is switched off. In the case of DKD
calibration of the temperature calibrators, the errors of
measurement of the reference are exactly determined and the
parameters P0 to P4 are set to the optimum.
Fine linearisation can be disabled by setting the switch to OFF.
This function simplifies the determination of new parameters Pn
for adjusting the reference, at an autuorised laboatory.
The user should not edit these settings.
3.1.3.1 Internal Reference
Fig. 23: Internal reference
parameter set
Summary of Contents for CTD9300 Series
Page 67: ...9037926 07 2015...
Page 68: ...9037926 07 2015...