Temperature Dry Well Calibrators
CTD9300
3. Moduses and Functions
GB
46
WIKA Operating Instructions
CTD9300
Rev. 1
Fig. 31: Calibration menu
“min/max”
In addition to the temperature
display, the min. and max.
values of the current block
temperature are displayed. The
two values are displayed
continuously. The registration of
the min. and max. values is
restarted with the ENTER key.
The variation limits of the
temperature in the calibrator can
easily be displayed through the
min. and max. temperatures.
3.3.1.1 Calibration Menu “min/max”
Fig. 32: Calibration menu
“standard deviation”
In addition to the temperature
display, the standard deviation
of the current block temperature
is continuously displayed. The
recalculation of the standard
deviation is restarted with the
ENTER key.
The stability of the temperature
in the calibrator can easily be
displayed through the standard
deviation.
3.3.1.2 Calibration Menu “sta. dev.”