6. Commissioning, operation
6.7.5 SETUP additional menu items: CPH configuration
This menu section lists general instrument settings, such as:
1
Input possibility for an ambient temperature
2
Input possibility for a height difference between reference pressure sensor and test
item, used in the automatic correction (deduction of a medium column)
The standard height difference between test item and CPT6000 reference
pressure sensor is 98 mm (3.86 in), see Fig. 1 “Dimensioning of the height
difference” in chapter 6.2.1 “Compensation for height differences”).
3
Selection of the menu language (German/English/French/Spanish/Italian)
4
Date of the system clock (year/month/day)
5
Time of the system clock (Hours/Minutes/Seconds)
6
Brightness of the backlighting of the display
7
Powersave function (automatic switch-off time for backlighting and internal power
supply for test items with DC 24 V).
If the instrument is idle for the set switch-off time (no buttons pressed and no
interface communication), then the backlighting and the voltage supply for the test
item with DC 24 V will be switched off until any button is pressed or the instrument is
addressed over the interface.
8
Current rechargeable battery capacity
At 10 % the low battery charge warning, “low BAT” is shown in the display.
1
2
3
4
5
6
7
8
Input of current term
100 %
Batterylevel:
15 min
Powersave:
80 %
Brightness:
13:22:51
Time:
2011/05/10
Date:
German
Language:
0 mm
HightDiff:
°C
20.0
Temperature:
..CPH-Configuration
Selection of menu item
Input via numeric keypad
Confirmation of input with
[ENTER]
(
[CLEAR]
deletes the input)
ENTER