WIKA operating instructions, model CPH7650
EN
14261987.02 01/2019 EN/DE
6. Commissioning, operation
Operating mode: CALIBRATION
In the
CALIBRATION
mode the data shown above the dashed dividing line is the same
as in the “
MEASURING
with test item” mode.
Display indication
CALIBRATE
with test item
1
Current measured value of the CPT6000 reference pressure sensor
2
Current measured value of the test item
3
Deviation between the test item and reference
4
Set point of the calibration
5
Actual value of the calibration
6
P - 01: Test item No. 1
<01>: Test step No. 1
7
IDENT number of the test item
8
Current value of the output signal of the test item
P
I/U
switchable
123456
ID no:
01
P-01:
bar
0.000
True:
bar
0.000
Set:
%FS
0.07
bar
0.004
Dev.:
bar
4.011
4-20 mA
Testitem
bar
0.000
10.000
0.000
Reference
13:41
Calibration
8
Electrical signal (test
item)
3
4
5
6
7
1
3
123456
01
bar
0.000
bar
0.000
%FS
0.07
bar
0.004
bar
0.004
6.000
0.000
bar
0.000
10.000
0.000
Reference
Calibration
13:41
P-01:
Set:
True:
ID no:
Dev.:
Testitem
Pressure signal (test
item)