6. Commissioning, operation
Menu Param.
Setting
Meaning
´MENU´
▶
▲
or
▼
SEt
ConF
Set Configuration: Generic settings
Unit
mbar, bar, ...
Unit: Unit of display
*, **
SL
oFF / on
Sea-level correction: on or off
*, **
Alti
-200 ... +9999
Altitude: Input of altitude above sea-level [m] (only if on)
*, **
rAtE
Rate: Measuring rate (see chapter 6.3.3)
*
Slo
Slow: Measuring rate (4 Hz filtered, low power consumption)
*
FASt
Fast: Measuring rate, filtered (> 1000 Hz)
*
P.dEt
Peak detection: fast measuring rate, unfildered (> 1000 Hz)
*
t.AVG
1 ... 120
Averaging interval in seconds, used by the averaging function off
oFF
Averaging function deactivated
P.oFF
1 ... 120
Auto Power-Off delay in minutes. Device will automatically switch
itself off if, during this interval, no key is pressed, or there is no
serial communication.
oFF
Auto Power-Off function inactive (continuous operation)
Out
oFF
Function of the output: No output function, lowest power consump-
tion
SEr
Output is serial interface
dAC
Output is analogue output 0 ... 1 V
Adr.
01, 11 ... 91
Communications address of the interface
dAC.
CH1, CH2 or
CH DIF
Choice of the input to be the source for the analogue output (if Out
= dAC)
dAC.0
eg. -5.00 ...
+5.00 mbar
Enter desired value at which the analogue output potential should
be 0 V (if Out = dAC)
dAC.1
eg. -5.00 ...
+5.00 mbar
Enter desired value at which the analogue output potential should
be 1 V (if Out = dAC)
SEt
CAL
Set Calibration: Adjustment of sensors
OFS.1
Sensordep.,
e.g. -5.00 ...
+5.00 mbar
The
offset of sensor 1
will be displaced by this value to compen-
sate for deviations in the probe or in the measuring device.
OFF
Zero displacement inactive (=0.0 °)
SCL.1
eg. -5.00 ...
+5.00 mbar
The measuring
scale of sensor 1
will be changed by this factor
[%] to compensate deviations of probe or measuring device.
OFF
Scale correction factor inactive (=0.000)
OFS.2
Sensordep.,
e.g. -5.00 ...
+5.00 mbar
The
offset of sensor 2
will be displaced by this value to compen-
sate for deviations in the probe or in the measuring device.
OFF
Zero dispacement deactive (=0.0°)
SCL.2
eg. -5.00 ...
+5.00 mbar
The measuring
scale of sensor 2
will be changed by this factor
[%] to compensate deviations of probe or measuring device.
OFF
Scale correction factor inactive (=0.000)
Summary of Contents for CPH62I0-S1
Page 44: ...44 WIKA operating instructions model CPH62I0 11221801 02 02 2011 GB D F E GB...
Page 85: ...85 WIKA Betriebsanleitung Typ CPH62I0 D 11221801 02 02 2011 GB D F E...
Page 86: ...86 WIKA Betriebsanleitung Typ CPH62I0 D 11221801 02 02 2011 GB D F E...
Page 122: ...122 WIKA mode d emploi type CPH62I0 F 11221801 02 02 2011 GB D F E...