1. Setup
2. Program operation and procedures
2.1 Calibration Procedures
1. Connect EMM to sled as shown on page 3. Apply power.
2. Start EMM Calibration program. It should appear as shown in Figure 2.1.
Figure 2.1 – Main screen
3. Set Com Port to your computers com port value. The baud rate settings are built into the application. They are 57600, N, 8, 1.
The Com Test button brings up the window below. If the EMM is not connected or a problem exists an error will be shown.
Figure 2.2 A – Successful com test
Figure 2.2 B - Communication error
1.2 Software installation
To install the software, run the setup program on the included disk (setup.exe). The setup program will install all the necessary
files to run the program. It will register all files appropriately so it can be completely removed using the Windows® add/remove
program. Upon installation, a desktop shortcut and start menu shortcut will be added. If this is not necessary, the “Xducer_Utility.
exe” program may be run directly.
The computer on which the software is run must have Windows® XP or greater. It must have an available RS-232 serial port
capable of running at 57600 baud. Finally, Microsoft .NET Framework 4.0 or greater must be installed. This is included as the file
“dotNetFx40_Full_x86_x64.exe”. The latest version can be downloaded from: http://www.microsoft.com/net
Calibration Sled Hardware Setup: Connect the AC adapter of the power supply to an AC power socket. Then connect the power
supply to the circular power connector on the rear of the calibration sled. Connect the 9 pin d-sub connector on the calibration sled
to any available RS-232 connector on the computer. This connection is made through the straight 9 pin male to 9 pin male cable.
Transducer Com Cable Setup: Connect the AC adapter of the power supply to an AC power socket. Then connect the “Computer”
side of the cable to any available RS-232 connector on the computer. Connect the “Transducer” side of the cable to the EMM
communication port.