background image

5

WIKA operating instructions EMM Calibration Sled 

PN 0019408001A  08/2018  EN

EN

3. Program operations and procedures

4.  Select Calibrate button. The next screen should appear as shown in Figure 2.3. 

Figure 2.3 – Calibration/data screen

The following fields may be updated:

• Calibration Date – a six digit number must be entered.  The calibration date entered should be formatted in the form “MMDDYY” 

or “DDMMYY” so that the Mensor instrument will convert the date correctly for the different locales available from the supported 

languages.
• Cert # – a 16 character string can be entered.  The Cal ID# is an arbitrary string you may use to identify the calibration of the 

sensor.
• Units  – Displays the native units of the sensor
• Zero  – a value between ± 0.1% of range, usually close to 0. This is the calibration offset used to adjust the zero reading of the 

sensor. The zero offset is unit based so enter the value in the native units of the sensor.
• Span  – a value between 1 ± 0.15% of range, usually close to 1.0. This is the calibration correction multiplier used to adjust the 

span reading of the sensor.

5. The tool is divided into four separate sections corresponding to each sensor in the EMM. 

  • Head temperature is the Platinum Resistive Thermometer (PRT) for head temperature measurement that has an external probe. 

  • Ambient Temperature is the internal ambient temperature sensor. 

  • Ambient Pressure is the internal barometric pressure sensor. 

  • Relative Humidity is the internal humidity sensor. 

Observe each reading to determine if sensors are live.

6.  The PRT should be calibrated by probe immersion in a precision temperature bath against a standard that has accuracy as 

previously described. Multiple choices exsist for calibration. Including forced zero and/or span adjustment with the same named 

buttons. The 2-pt Cal is a way to calibrate at two discrete tempertures. Linearity is a way to calibrate multiple discrete temperture 

points and auto-linearizing the PRT.

7. All other sensors can be calibrated using the Zero, Span or 2-pt Cal buttons. 

 

Considering the accuracy requirements of these sensors, at Mensor we find that a simple single point calibration using the 

Zero button under ambient conditions is sufficient.

8. If pressure is to be applied to the barometric sensor for calibration, tubing and an adaptor fitting have been provided. One end 

of the tubing is inserted into the hole in the EMM’s case. The tubing can then be applied over the sensor barb. The other end of the 

tubing can be connected to a pressure source with the adaptor fitting. The adaptor fitting has a 7/16-20 male thread

Head temperature linearity

Selection of the linearity button brings up the following screen. 

Summary of Contents for CPD8500

Page 1: ...s reserved Mensor is a registered trademark of Mensor LP All other brand and product names are trademarks or registered trademarks of their respective companies Prior to starting any work read the operating instructions Keep for later use Transducer utility manual 0020646001 ...

Page 2: ...A 08 2018 EN EN Contents Contents 1 Introduction 3 1 1 Equipment required 3 1 2 Technincal information 3 2 Program operation and procedures 4 2 1 Calibration Procedures 4 2 2 Software installation 4 2 3 Use of multiple calibration methods 7 2 4 Additional features 8 ...

Page 3: ...n platinum resistive thermometer PRT The rated accuracies are Initial Accuracy Stability Ambient Temperature 4 C 2 C year Relative Humidity 3 2 year Barometric Pressure 25 2 year PRT 2 C 1 C year It is important to use standards that have a test uncertainity ratio of at least 1 1 A higher ratio is better 1 3 Equipment required EMM Calibration Sled P N 0019780001 EMM P N 0019582001 CPD8500 EMM Cal ...

Page 4: ...a desktop shortcut and start menu shortcut will be added If this is not necessary the Xducer_Utility exe program may be run directly The computer on which the software is run must have Windows XP or greater It must have an available RS 232 serial port capable of running at 57600 baud Finally Microsoft NET Framework 4 0 or greater must be installed This is included as the file dotNetFx40_Full_x86_x...

Page 5: ...head temperature measurement that has an external probe Ambient Temperature is the internal ambient temperature sensor Ambient Pressure is the internal barometric pressure sensor Relative Humidity is the internal humidity sensor Observe each reading to determine if sensors are live 6 The PRT should be calibrated by probe immersion in a precision temperature bath against a standard that has accurac...

Page 6: ... entering the reference value then selecting the Take Reading button A graph is updated as the points are entered When the table is fully populated select Calibration and Save Calibration The linearity correction will be applied and saved Figure 2 5 Linearity correction graph Other options are available such as reducing the number of calibration points and upscale downscale calibrations By changin...

Page 7: ...e used to change the slope This value is nominally 1 This is a multiplier such that 1 would not change the current slope A value of 0 9 would multiply the current slope by 0 9 or 90 The Zero and Span window is shown in Figure 4 Figure 2 6 Zero and Span window 2 pt cal Use of the 2 pt Cal button requires the capability to apply two discrete calibration stimuli to the individual sensors A reference ...

Page 8: ...A Alexander Wiegand SE Co KG Alexander Wiegand Straße 30 63911 Klingenberg Germany Tel 49 9372 132 0 Fax 49 9372 132 406 info wika de www wika de Mensor Corporation 201 Barnes Drive San Marcos TX 78666 USA Tel 1 512 3964200 15 Fax 1 512 3961820 E Mail sales mensor com www mensor com ...

Reviews: