7

WIKA operating instructions EMM Calibration Sled 

PN 0019408001A  08/2018  EN

EN

                 

   

 

Figure 3.9 A - 5 points selected   

Figure 3.9 B - Reading upscale and downscale

Figure 3.9 C - Reference and reading upscale and downscale

 

Note: At any point these processes can be aborted and the Factory Restore button can be selected. This will revert the 

calibration back to the factory settings.

2.2  Use of multiple calibration methods

 

Note: All entries should be followed by a carriage return for values to be accepted.  

Zero or Span buttons: 

 

Zero can be used to force an offset in the value shown by entering a value in the same units for example °C. 

 

THE ZERO BUTTON SHOULD BE USED FOR SINGLE POINT CALIBRATIONS.

 

Span can be used to change the slope. This value is nominally 1. This is a multiplier such that 1 would not change the current slope. A 

value of 0.9 would multiply the current slope by 0.9 or 90%. The Zero and Span window is shown in Figure 4.

Figure 2.6 - Zero and Span window

2-pt cal:

Use of the 2-pt Cal button requires the capability to apply two discrete calibration stimuli to the individual sensors.  

A reference value near the zero point of the sensor should be entered. When the Enter button is selected the application will read 

the sensor’s value and automatically make the correction. The same applies to the next point. It is important that the reference 

stimulus be close to the “Apply” value shown. The 2-pt Cal windows are shown below.

   

Figure 2.7 A - Displays for both temperatures sensors 

   

Figure 2.7 B - Displays for both barometric pressure sensors

  

Figure 2.7 C - Displays for both relative humidity sensors

Summary of Contents for CPD8500

Page 1: ...s reserved Mensor is a registered trademark of Mensor LP All other brand and product names are trademarks or registered trademarks of their respective companies Prior to starting any work read the operating instructions Keep for later use Transducer utility manual 0020646001 ...

Page 2: ...A 08 2018 EN EN Contents Contents 1 Introduction 3 1 1 Equipment required 3 1 2 Technincal information 3 2 Program operation and procedures 4 2 1 Calibration Procedures 4 2 2 Software installation 4 2 3 Use of multiple calibration methods 7 2 4 Additional features 8 ...

Page 3: ...n platinum resistive thermometer PRT The rated accuracies are Initial Accuracy Stability Ambient Temperature 4 C 2 C year Relative Humidity 3 2 year Barometric Pressure 25 2 year PRT 2 C 1 C year It is important to use standards that have a test uncertainity ratio of at least 1 1 A higher ratio is better 1 3 Equipment required EMM Calibration Sled P N 0019780001 EMM P N 0019582001 CPD8500 EMM Cal ...

Page 4: ...a desktop shortcut and start menu shortcut will be added If this is not necessary the Xducer_Utility exe program may be run directly The computer on which the software is run must have Windows XP or greater It must have an available RS 232 serial port capable of running at 57600 baud Finally Microsoft NET Framework 4 0 or greater must be installed This is included as the file dotNetFx40_Full_x86_x...

Page 5: ...head temperature measurement that has an external probe Ambient Temperature is the internal ambient temperature sensor Ambient Pressure is the internal barometric pressure sensor Relative Humidity is the internal humidity sensor Observe each reading to determine if sensors are live 6 The PRT should be calibrated by probe immersion in a precision temperature bath against a standard that has accurac...

Page 6: ... entering the reference value then selecting the Take Reading button A graph is updated as the points are entered When the table is fully populated select Calibration and Save Calibration The linearity correction will be applied and saved Figure 2 5 Linearity correction graph Other options are available such as reducing the number of calibration points and upscale downscale calibrations By changin...

Page 7: ...e used to change the slope This value is nominally 1 This is a multiplier such that 1 would not change the current slope A value of 0 9 would multiply the current slope by 0 9 or 90 The Zero and Span window is shown in Figure 4 Figure 2 6 Zero and Span window 2 pt cal Use of the 2 pt Cal button requires the capability to apply two discrete calibration stimuli to the individual sensors A reference ...

Page 8: ...A Alexander Wiegand SE Co KG Alexander Wiegand Straße 30 63911 Klingenberg Germany Tel 49 9372 132 0 Fax 49 9372 132 406 info wika de www wika de Mensor Corporation 201 Barnes Drive San Marcos TX 78666 USA Tel 1 512 3964200 15 Fax 1 512 3961820 E Mail sales mensor com www mensor com ...

Reviews: