button ,observe the value change on the display , release the button when fluctuation
is less or equal to ±1.
5. Press the key on the operation panel, complete calibration process.
Caution
Before the calibration, please make sure you can provide two
temperature standard points between 20 to 200
℃
, and the
difference between the two points should be larger than 50
℃
.
4.3.5. Using the RS232 Connection
The hot plate provides a RS-232 connection for temperature monitoring. The actual
temperature data will be sent through the RS-232 interface every 3 seconds. The data
can be read with any software which can acquire data with from an RS-232 connection.