Installation, operating and maintenance manual
UNIGAS 300
DDG6004MHGB/08-2017/Rev.A2 Software version from M 1.3.40 and D 2.3.37
37
Objects/parameters that define the conversion, inputs and correction of the gas meter measuring error
Protection level: calibration lock
Name
Unit
Menu
item
OBIS code
Description
INP1DIV
imp/m
3
7-1
7-1:0.7.2
Scaling factor for input 1
INPHFIDIV imp/m³
7-1
7-4:0.7.2
Scaling factor for input 1 if this is set for HF impulses
INP2DIV
imp/m
3
7-1
7-2:0.7.2
Scaling factor for input 2
INP3DIV
imp/m
3
7-1
7-3:0.7.2
Scaling factor for input 3
Input 1
-
C.93.9
Functionality of input 1: LF, HF, encoder and encoder at
Vm2
For
Encoder at Vm2,
Vm1 counts LF impulses presented
at LF1 input. In Vm2 the encoder reading is counted on
the basis of 5-min interval reading
Cf1
-
C.98.1
Factor 1 for measuring error correction of gas meter.
Between Qf1 and Qf2 linear interpolation takes place
between Cf1 and Cf2
Qf1
m
3
/h
C.98.2
Qf1 where Qf1= Q
min
gas meter (see EN12405-1 +A2)
Cf2
-
C.98.3
Factor 2 for measuring error correction of gas meter.
Between Qf2 and Qf3 linear interpolation takes place
between Cf2 and Cf3
Qf2
m
3
/h
C.98.4
Qf2 where Qf1 < Qf2 < Qf3
Cf3
-
C.98.5
Factor 3 for measuring error correction of gas meter.
Between Qf3 and Qf4 linear interpolation takes place
between Cf3 and Cf4
Qf3
m
3
/h
C.98.6
Qf3 where Qf2 < Qf3 < Qf4
Cf4
-
C.98.7
Factor 4 for measuring error correction of gas meter.
Between Qf4 and Qf5 linear interpolation takes place
between Cf4 and Cf5
Qf4
m
3
/h
C.98.8
Qf4 where Qf3 < Qf4 < Qf5
Cf5
-
C.98.9
Factor 5 for measuring error correction of gas meter.
Outside the range of Qf5 correction takes place with Cf5
Qf5
m
3
/h
C.98.10
Qf5 where Qf4 < Qf5 < Q 6
Cf6
-
C.98.11
Factor 6 for measuring error correction of gas meter.
Between Qf6 and Qf7 linear interpolation takes place
between Cf6 and Cf7
Qf6
m
3
/h
C.98.12
Qf6 where Qf5 < Qf6 < Qf7
Cf7
-
C.98.13
Factor 7 for measuring error correction of gas meter.
Between Qf7 and Qf8 linear interpolation takes place
between Cf7 and Cf8
Qf7
m
3
/h
C.98.14
Qf7 where Qf6 < Qf7 < Qf8
Cf8
-
C.98.15
Factor 8 for measuring error correction of gas meter.
Between Qf8 and Qf9 linear interpolation takes place
between Cf8 and Cf9
Qf8
m
3
/h
C.98.16
Qf8 where Qf7 < Qf8 < Qf9
Cf9
-
C.98.17
Factor 9 for measuring error correction of gas meter.
Between Qf9 and Qf10 linear interpolation takes place
between Cf9 and Cf10
Qf9
m
3
/h
C.98.18
Qf9 where Qf8 < Qf9 < Qf10
Cf10
-
C.98.19
Factor 10 for measuring error correction of gas meter. If
Q > Q10 is corrected with value Qf10
Qf10
m
3
/h
C.98.20
Qf10 where Qf10= Q
min
gas meter (see EN12405-1 +A2)