CL24s
USER’S MANUAL
4. Recommended separation distances between portable and mobile RF
communications equipment and this medical LCD monitor.
The medical LCD monitor is intended for use in an electromagnetic environment
in which radiated RF disturbances are controlled. The customer or the user of the
monitor can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters)
and the medical LCD monitor as recommended below, according to the maximum
output power of the communications equipment.
Rated maximum
output power [W]
of transmitter
Separation distance according to frequency of transmitter[m]
150kHz to 80MHz
d = [ 3.5 / V
1
] √P
80MHz to 800MHz
d = [ 3.5 / E
1
] √P
800MHz to 2.5GHz
d = [ 7 / E
1
] √P
V
1
=3Vrms
E
1
=3V / m
E
1
=3V / m
0.01
0.116
0.116
0.2333
0.1
0.368
0.3687
0.7378
1
1.166
1.1660
0.2333
10
3.687
3.6872
0.7375
100
11.660
11.6600
23.333
For transmitters rated at a maximum output power not listed above, the recommended separation
distance(d) in meters (m) can be estimated using the equation applicable to the frequency of
the transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
Note
1. At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
2. These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Summary of Contents for CL24s
Page 39: ...1 2 3 5 7 FCC 10 15 17 18 19 OSD 19 OSD 24 29 30 32 33 33...
Page 40: ...CL24s CL24s 24 LCD 10 bit CL24 CL24 IPS LCD LCD IPS CL24s...
Page 41: ...3 DC 6...
Page 42: ...CL24s UL 60601 1 CAN CSA C22 2 No 601 1 FCC B CCC RoHS...
Page 45: ...7 y y 0 40 20 60 y y 30 80 y y 500 1060 hPa LCD...
Page 47: ...9 LCD LCD y y y y LCD y y LCD y y LCD y y LCD y y y y LCD y y LCD y y LCD...
Page 48: ...CL24s LCD y y I AC DC y y y y y y FCC LCD FCC 15 B 1 2 LCD 3 4...
Page 54: ...CL24s 24 1 CL24s FS L2403D 24 1 CL24s FS L2403D...
Page 55: ...17 0 108mm 108mm 30 30 30 30 3 30 30 3 y y y...
Page 57: ...19 VESA DPMS LED CL24s FS L2403D 60 W 10 W 1 W OSD FSN OSD OSD OSD 1 OSD OSD OSD...
Page 58: ...CL24s 2 OSD OSD 3D PIP...
Page 63: ...25 D SUB C1 1 MODE A C1 REC709 D65 C2 2 C3 3 BYPASS 16 9 1 1 16 9 1 1...