28
Pro 200 Series
Furnace Operations Manual
Fuses
The furnace contains two fuses:
• On the rear, just below and to the right of the power
cord, there are two 12 Amp (115 Vac) Slo-Blo or two
7 Amp (220 Vac) Slo-Blo, .25 x 1.25 inch ceramic
fuses.
If More Help is Needed
We hope you have many years of trouble-free service from
your furnace. If you do have problems with the furnace, or
if you have questions about the furnace not covered in
the manual, contact your dealer or Whip Mix at:
www.whipmix.com
Phone: 800-626-5651
Fax: 502-634-4512
E-Mail: [email protected]
Be prepared to provide the following information:
1. Your name
2. Your lab’s name and address
3. Your lab’s phone number
4. Your lab’s fax number
5. Furnace model and serial number (serial number
can be found on the rear of the furnace)
6. Your question/problem
When you call, it would be helpful if you are near the
furnace. The technician will probably ask you to run tests
and report the results, or read the display while the test is
running.
Vacuum Test
1. Press the
Up
arrow button to raise the table fully up.
You cannot start a vacuum test until the table is fully up.
2. Select the
Special Functions
item.
3. Turn the
Selector Knob
to
Test Furnace
. Press the
Next
button.
4. Turn the
Selector Knob
to
Vacuum Test
. Press the
Next
button. The vacuum test will automatically begin.
There will be three, numerical indicators on the screen,
each has its own individual meaning.
• The number to the far left is the vacuum count. That
is the vacuum sensor reading.
• The second in the middle is the centimeters or inches
of mercury.
• The third is the current condition of the table
0 = moving, 99 = fully up or fully down.
5. Watch the numerical indicators as they rise.
6. The third value should stay at 99 and not change.
Sometimes the value changes from 99 to 0 when the
vacuum pump starts. This is normal and happens on
some units.
7. Once the vacuum has reached 240, press the
Next
button. This will place the test in the hold position
and allow you to watch for changes in vacuum value
and readings.
The acceptable loss is 2 cm within six minutes.
8. Once the test has been in hold for 6 minutes, press
the
Next
button and the test will end.
9. If the numbers dropped outside the acceptable loss
range, then the vacuum system has a leak and it must
be repaired and recalibrated.
Test Furnace
> Vacuum Test
Press Test
Muffle Test
*
Special Functions
Export Settings
Import Settings
> Test Furnace
Vacuum Test On
90
00CM
99
Next=Test Hold
Vacuum Test On
223 65CM
99
Next=Test Hold