Single Channel EEG Test System | User Manual
9
․
When the USB module is connected for the first time, select manual
installation, and point to the folder containing the above file. Then
continue to follow the instructions to finish the installation. There may be
a warning that the driver is not recognized by Windows®, and this can be
ignored. Please click
to watch the tutorial video.
2.1.4 First Time Using WhaleTeq Product
–
Microsoft .Net Framework
4.0 Installation
WhaleTeq software is developed by Microsoft .Net Framework 4.0. If SEEG
software fails to launch properly, please check whether Microsoft .Net
Framework 4.0 or higher versions was installed in the operation system.
If your PC does not install Microsoft .Net Framework 4.0 or higher versions,
please download from Microsoft website. Please click here to watch the
tutorial video (from 2:03).
2.2 Connecting to the EEG
For connecting the EEG device to the SEEG 100E and use the provided
22 channels “EEG
breakout box”.
There are only 9 terminals (Ch1~Ch9) could output waveforms, and the rest
13 terminals (Ch10~Ch22) are connected to ground. Below is the photo of EEG breakout
box.
Alternately the EEG device under test can be directly connected to the SEEG 100E module
using a male D15 connector. The pin outs are:
1-Ch1
4
–
GND
7
–
Ch6
10
–
Ch9
2-Ch2
5
–
Ch4
8
–
Ch7
3-Ch3
6
–
Ch5
9
–
Ch8
Note: Ch10 ~ Ch22 connect to pin 4 (GND).