Single Channel EEG Test System | User Manual
16
Load waveform:
The
“Load waveform” function supports
three formats
–
Text, Binary and EDF
files.
Text (*.txt)
-
Ascii file, Windows line breaks (LF, CF)
-
first line is sample rate (Hz)
-
second line number of samples
-
following lines are samples in microvolts (one sample per line)
Binary files (*.bin)
-
Bytes 1-2 are sample rate (Hz)
-
Bytes 3-6 are number of samples
-
Following bytes are samples, 2 bytes per sample
-
all data is big-endian (high byte first), 2's compliment
EDF files (*.edf)
This is a commonly used but complicated format.
for the format
details.
2.6 IEC 80601-2-26 Helper
It is a companion software add-on to enhance the function of SEEG 100E. It supports the
latest EEG standards IEC80601-2-26:2019 with detailed preset parameter settings and
actual test sequence for testing needs.