Single Channel EEG Test System | User Manual
4
1.2 Standards/Application
The following table shows the standards for which this system has been designed
for, and includes any limitations:
Standard
Clause(s)
Limitations/Notes
IEC 80601-2-
26:2012
201.12.1
201.12.1 all performance tests except
201.12.1.104 input noise and
201.12.1.106 CMRR tests
General limitation:
(1) This equipment is designed for use with isolated EEG circuits, as are generally
provided for medical EEG. If applied to a non-isolated circuit, the noise may be
excessive.
(2) In EEG breakout box, there are 22 terminals. But there are only 9 terminals
could output waveforms, the rest 13 terminals are connected to ground.