76
AECG100 | User Manual
Parameter
Specification
(variable, may include up to
50
μ
Vpp noise)
Accuracy
±
1%
RCA output amplitude
Range
ECG electrode output x1000,
max 10V
Accuracy
±1% for amplitudes of 0.5Vpp
or higher
9.2
PPG Test Mode
Table 5: PPG Test Mode Specification
Parameter
Specification
BPM
Range
10~300 BPM
Resolution 1 BPM
Accuracy
±1 BPM
LED AC Level reference
output (RCA)
Range
0.075~3 V (Full range)
Resolution (1/3000) full range
Accuracy
±
1%
Lumen
§
Range
50 ~ 925 Lux (full range)
(5mm from diffusion film)
Resolution (1/3000) full range
Accuracy
±
6%
Light wavelength
Range
525nm
Accuracy
±
10nm
S/N Ratio
>51dB (with USB isolator)
PD response time
Rising
1
μ
s Typical*
Falling
1
μ
s Typical*
§
The data is for PPG-1R-525 module.
* The PD response time will vary with the light intensity of the DUT.