![WHALETEQ AECG100 User Manual Download Page 26](http://html.mh-extra.com/html/whaleteq/aecg100/aecg100_user-manual_967648026.webp)
26
AECG100 | User Manual
Software Interface
4.1
General
Figure 19: Button Functions in General Mode
(1)
PPG Config.:
This function belongs to the PPG module and configures PWTT, SpO
2
,
PPG, and Auto Sequence (where applicable) tabs. Different DUTs might
have different signal processing, so AECG100 provides the
“I
nverted
”
function to output the inverted waveform, which can be confirmed on
an oscilloscope. Check
“
Synchronized Pulse
”
to enable the
synchronization behavior between AECG100 LED and DUT LED.
Figure 20: PPG Config.
Trigger level: Set the trigger level that drives the AECG100 LED to turn
on after the AECG100 PD detects the value of the LED brightness of the
DUT.
Ambient light: Available when connecting the transmission SpO
2
module PPG-2TF-660 to the AECG100 main console to add 14 types of
simulated indoor/outdoor ambient light signals to the output
waveform.
1
2