47
AECG100 | User Manual
or 10 sec time frames.
(9)
Load:
It is required to load PPG/ ECG raw data separately by
different
“C
hannels
”
(ECG/ PPG Green/ PPG RED/ PPG Infrared).
Either saved from (1) Signal tab “Waveform Raw” or self
-
constructed according to WhaleTeq rule (click
“
Help
”
for more
details), t
he raw data file can be loaded by clicking the “File”
section.
The “Signal” number corresponds to channel types.
The
“
Total Length
”
displays the total playback time. Users have to
manually adjust “Output Setting” to ensure AECG100 output
s
signal properly
.
Figure 53: Waveform Player Setting
(10)
Play/ Stop:
Play the selected raw data file in *.txt
format. Once
stopped, play from the beginning.
(11)
Loop:
Check to replay the raw data file repeatedly.
(12)
Current Position:
Display the timing of the played raw data.