56
AECG100 | User Manual
desired SpO
2
level (%) without modifying the intercept and slope,
the R value will be generated in accordance with R-Curve equation.
The same R Value is used to display the corresponding
RED/Infrared parameters.
For unknown R curve, users may obtain the intercept and slope of
the R-Curve equation by using the SpO
2
Table (Please refer to the
description of (12) Table).
Figure 65: Setting R Curve
(4)
Output Waveform Selection:
Select to output different
waveform types (ex. Sine, Triangle, or PPG).
(5)
Use SpO
2
Table:
Option 1. User can
click “Table”
to build SpO
2
table with
specified parameters of R/IR LED.
Figure 66: Using SpO
2
Table
After completing the SpO
2
table, click
“
Graph
”
to obtain the R
curve equation.
Intercept
Slope
RED/Infrared LED Parameter Setting
R Curve
12