10

 

USER ACTION

Open the lid of centrifuge, correct the unbalanced load 

by various methods described in the balancing the rotor 

part of this manual. Then close the lid and `reset the 

centrifuge’. Resetting is done by switching `Off’ and 

again switching `On’ the centrifuge by using the On/Off 

switch located on the rear side of the centrifuge.

15. WARRANTY STATEMENT

14. CARE & MAINTENANCE

This personal centrifuge and its rotor should be cleaned 

periodically. Ensure that while cleaning the unit, it is not 

plugged in to the main power supply. To clean this 

centrifuge, use a damp cloth and a mild, noncorrosive 

detergent (ph < 8). Do not immerse the centrifuge in 

liquid or pour liquid over it. Excessive amounts of liquid 

should be avoided, liquid should not come into contact 

with the motor. After cleaning, ensure all parts are dried 

thoroughly before attempting to operate the unit.

Should a spill of infectious materials occur within the 

rotor or rotor chamber, the unit should be disinfected. 

This should be performed by qualified personnel with 

proper protective equipment. Do not use any cleaning 

or decontamination methods except those recommended 

by the manufacturer. Check with the manufacturer that 

the proposed method will not damage the equipment.

This product is warranted to be free from defects in 

material and workmanship for a period of two (2) years 

from date of purchase. Your product will be duly 

repaired upon prompt notification in compliance with 

the following conditions :
This warranty is valid only if the product is used for its 

intended purpose and within the guidelines specified 

in this instruction manual. This warranty does not cover 

damage caused by accident, neglect, misuse, improper 

service, natural forces or other causes not arising from 

defects in original material or workmanship. This 

warranty does not cover any incidental or consequential 

damages, commercial loss or any other damages from 

the use of this product.

WESTLAB
SMART MICRO
CENTRIFUGE

USER MANUAL

CODE  663-958

C

Summary of Contents for 663-958C

Page 1: ...nical Specifications 2 5 Standard Accessories 2 6 Safety Precautions 2 7 Installation 3 8 Standard Part Listing 4 9 Rotor Balancing 4 10 Interchageable Rotor 5 11 Changing the Rotor 5 12 Opening Centr...

Page 2: ...e on lid opening for user safety Quick start stop for short spin Smooth and quiet running Tiny footprint It can squeeze easily on to a crowded counter top Interchangeable PCR rotor adaptor Low voltage...

Page 3: ...4 C to 45 C Main supply voltage fluctuations not to exceed 10 of the nominal voltage Position at least 20 cm from walls 20 cm from ceiling and 20 cm between other products Make sure to always install...

Page 4: ...ied micro centrifuge tubes made from plastic and designed to withstand centrifugal forces of at least 2000 g Do not operate the centrifuge without the rotor being properly attached to the shaft Re che...

Page 5: ...e serious damage to the centrifuge 8 STANDARD PARTS LISTING ON OFF SWITCH NON SLIP RUBBER BASE POWER SOCKET PROTECTIVE LID ROTOR SAFETY SWITCH DISPLAY RELEASE BUTTON SPEED KNOB x 100RPM Min 60 TIMER K...

Page 6: ...is located at the bottom of the rotor on the marked side To remove the rotor turn counter clockwise to loosen the screw and then gently pull the rotor up vertically Insert the new rotor from the top...

Page 7: ...ter clockwise direction Trying to open the centrifuge lid by pulling the Lid Release button up is the wrong method of opening the lid Pulling the Lid Release button up will break the button or damage...

Page 8: ...id accidents and damage to the centrifuge 4 Close the lid flip ON the front switch to start centrifugation 5 Once the centrifugation is completed turn the front switch to Off to gradually stop the rot...

Page 9: ...nd rotor the actual RCF value will vary One can refer to the table mentioned in the section 7 to get the actual RCF of different rotors at certain given speed values 8 The unit is equipped with a 0 25...

Page 10: ...ge to the centrifuge This centrifuge has an unique patented feature to protect the centrifuge its user the surroundings against such as accidental run When such an event occurs the sensor build inside...

Page 11: ...us materials occur within the rotor or rotor chamber the unit should be disinfected This should be performed by qualified personnel with proper protective equipment Do not use any cleaning or decontam...

Page 12: ...turned for service should be sent with postage prepaid in the original packaging or other suitable packaging padded to avoid damage We will not be responsible for damage incurred by improper packaging...

Reviews: