9. ROTOR BALANCING

1. Always balance the rotor before centrifugation. The

following aresymmetrical loadings of centrifuge tubes

to the rotor.

2. Ensure that the samples in the tubes are of

equal volume.

3. Symmetrical loading of tubes in the rotor is necessary,

otherwise vibration or imbalance can occur which

can cause serious damage to the centrifuge.

8. STANDARD PARTS LISTING

ON/OFF SWITCH

NON SLIP

RUBBER BASE

POWER 

SOCKET

PROTECTIVE LID

ROTOR

SAFETY 

SWITCH

DISPLAY

RELEASE 

BUTTON

SPEED 

KNOB

x 1

00

R

PM

M

in

60

TIMER 

KNOB

142

0.2 ml

35.5

718

319
567

886

1276

2000

1000

4500

3000
4000

5000

6000

188

0.5 ml

47

953

424
753

1177

1695

208

1.5 ml

52

1053

468
832

1300

1872

217

2 ml

54

1098

488
868

1356

1952

208

0.2 ml

52

1050

467
830

1297

1868

RCF CHART

ROUND ROTOR

PCR STRIP

RPM

WESTLAB
SMART MICRO
CENTRIFUGE

USER MANUAL

CODE  663-958

C

Summary of Contents for 663-958C

Page 1: ...nical Specifications 2 5 Standard Accessories 2 6 Safety Precautions 2 7 Installation 3 8 Standard Part Listing 4 9 Rotor Balancing 4 10 Interchageable Rotor 5 11 Changing the Rotor 5 12 Opening Centr...

Page 2: ...e on lid opening for user safety Quick start stop for short spin Smooth and quiet running Tiny footprint It can squeeze easily on to a crowded counter top Interchangeable PCR rotor adaptor Low voltage...

Page 3: ...4 C to 45 C Main supply voltage fluctuations not to exceed 10 of the nominal voltage Position at least 20 cm from walls 20 cm from ceiling and 20 cm between other products Make sure to always install...

Page 4: ...ied micro centrifuge tubes made from plastic and designed to withstand centrifugal forces of at least 2000 g Do not operate the centrifuge without the rotor being properly attached to the shaft Re che...

Page 5: ...e serious damage to the centrifuge 8 STANDARD PARTS LISTING ON OFF SWITCH NON SLIP RUBBER BASE POWER SOCKET PROTECTIVE LID ROTOR SAFETY SWITCH DISPLAY RELEASE BUTTON SPEED KNOB x 100RPM Min 60 TIMER K...

Page 6: ...is located at the bottom of the rotor on the marked side To remove the rotor turn counter clockwise to loosen the screw and then gently pull the rotor up vertically Insert the new rotor from the top...

Page 7: ...ter clockwise direction Trying to open the centrifuge lid by pulling the Lid Release button up is the wrong method of opening the lid Pulling the Lid Release button up will break the button or damage...

Page 8: ...id accidents and damage to the centrifuge 4 Close the lid flip ON the front switch to start centrifugation 5 Once the centrifugation is completed turn the front switch to Off to gradually stop the rot...

Page 9: ...nd rotor the actual RCF value will vary One can refer to the table mentioned in the section 7 to get the actual RCF of different rotors at certain given speed values 8 The unit is equipped with a 0 25...

Page 10: ...ge to the centrifuge This centrifuge has an unique patented feature to protect the centrifuge its user the surroundings against such as accidental run When such an event occurs the sensor build inside...

Page 11: ...us materials occur within the rotor or rotor chamber the unit should be disinfected This should be performed by qualified personnel with proper protective equipment Do not use any cleaning or decontam...

Page 12: ...turned for service should be sent with postage prepaid in the original packaging or other suitable packaging padded to avoid damage We will not be responsible for damage incurred by improper packaging...

Reviews: