Emissions and immunity information
Electromagnetic emissions
The ProBP™ 2000 Digital Blood Pressure Device is intended for use in the electromagnetic environment specified below.
The customer or user of the ProBP™ 2000 Digital Blood Pressure Device should assure that it is used in such an
environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The ProBP™ 2000 Digital Blood Pressure Device uses RF energy only for its
internal function. Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The ProBP™ 2000 Digital Blood Pressure Device is suitable for use in all
establishments, other than domestic and those directly connected to the
public low-voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration – electromagnetic immunity
The ProBP™ 2000 Digital Blood Pressure Device is intended for use in the electromagnetic environment specified below.
The customer or the user of the ProBP™ 2000 Digital Blood Pressure Device should assure that it is used in such an
environment.
Immunity test
IEC 60601 test
level
Compliance level Electromagnetic environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
power supply lines:
±2 kV
input/output lines :
±1 kV
power supply lines:
±2 kV
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
line(s) to line(s): ±1
kV
line(s) to earth: ± 2
kV 100kHz repetition
frequency
line(s) to line(s): ±1
kV 100kHz repetition
frequency
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
0% 0.5 cycle
At 0°, 45°, 90°,135°,
180°, 225°, 270°
and 315°
0% 0.5 cycle
At 0°, 45°, 90°,135°,
180°, 225°, 270°
and 315°
Mains power quality should be that of a typical
commercial or hospital environment.
32 EMC guidance and manufacturer’s declarations
Welch Allyn ProBP™ 2000 Digital Blood Pressure Device
Summary of Contents for ProBP 2000
Page 1: ...Welch Allyn ProBP 2000 Digital Blood Pressure Device Directions for use Software Version 1 X...
Page 4: ...iv Contents Welch Allyn ProBP 2000 Digital Blood Pressure Device...
Page 20: ...16 Maintenance Welch Allyn ProBP 2000 Digital Blood Pressure Device...
Page 26: ...22 Specifications Welch Allyn ProBP 2000 Digital Blood Pressure Device...
Page 30: ...26 General radio compliance Welch Allyn ProBP 2000 Digital Blood Pressure Device...
Page 32: ...28 Warranty Welch Allyn ProBP 2000 Digital Blood Pressure Device...
Page 34: ...30 Approved accessories Welch Allyn ProBP 2000 Digital Blood Pressure Device...