56
Repair Procedures
Welch Allyn Micropaq Monitor
3.
Insert the display flex cable into the J4 connector.
4.
When placing the assembly into the case, verify that the flap on the antenna covers
the ECG connector. Failure to cover will cause radio interference problems.
5.
Place the main board into the front case.
6.
Insert the keyboard cable into the J3 connector and lock the latching piece.
Note
Tilt the assembly slightly toward the connectors, as shown, to allow insertion of
the display flex cable into the J4 connector.
Caution
When attaching the J3 connector, bend the flex cable toward the
connector in a smooth curve (shown below at left) to avoid cable damage.
J4
Antenna flap
ECG connector
Antenna flap covering the ECG connector
J3