46
Repair Procedures
Welch Allyn Micropaq Monitor
Closing the Case
1.
Insert the rubber gasket (part of the fastener kit) into the groove in the front case.
Begin inserting the gasket in the end locking section, located on the lower edge of the
case (as shown below).
2.
After fully inserting the gasket into the groove, trim it to the specified length, as
shown in the illustration of the end locking section on
page 46
.
3.
Attach the battery and verify that:
Caution
Do not stretch the gasket. Stretching the gasket will make reassembly
more difficult, and may result in a poor seal between the front and rear case.
Note
Both ends of the gasket should extend as far as possible into the end locking
section (refer to the drawing).
Note
Do not trim the gasket until it has been fully inserted in the groove, with the
overlap as shown above.
Note
Take special care when inserting the gasket into the curved section of the groove
near the connectors. If the gasket is stretched, it may pull out of the groove at this
point.
Curved section of gasket groove