MAINTENANCE
40
Cleaning and Disinfecting Accessories
ECG Cables
Approved Cleaning Agents
Enzymatic detergent such as ENZOL (US) or CIDEZYME (outside the US).
Distilled water.
Disinfectant solution (such as CIDEX OPA, or a 10% solution of household
bleach (5.25% sodium hypochlorite) in distilled water).
Soft, lint-free cloths and/or soft-bristled brushes.
Protective gloves and eyewear.
Procedure
1.
Remove the ECG cable from the recorder.
2.
Put on gloves and protective eyewear.
3.
Prepare the detergent according to the manufacturer's instructions, and also
the disinfectant solution, in separate containers.
4.
Apply detergent to product using a soft, lint-free cloth. If material is dried on,
allow to sit for 1 minute. Do not immerse cable ends or lead wires in liquid as
it can cause corrosion.
5.
Wipe smooth surfaces with the cloth.
6.
Use a soft-bristle brush on visibly soiled areas and irregular surfaces.
7.
Remove detergent from product using cloth dampened in distilled water.
8.
Repeat as necessary.
9.
Apply disinfectant solution on affected area using a soft cloth. Allow product to
sit for 5 minutes.
10. Wipe excess solution and clean product again with cloth dampened in distilled
water.
11. Allow 2 hours for drying.
Carry case:
Wash the carry case by hand with fabric detergent and then air dry.
Do not machine dry the case.
The H12+ is compatible with a number of other accessories, each with unique cleaning and
disinfecting needs. Follow the cleaning and disinfecting instructions provided in the directions
for use shipped with those items.
WARNING: Always clean and/or disinfect accessories between patients to reduce the risk
of spreading infectious agents between persons.
WARNING: Do not reuse accessories indicated as single-patient use; as this may facilitate the
spread of infectious agents between persons.
WARNING: Prevent liquid from penetrating the device and do not attempt to clean/disinfect the
device or patient cables by submerging into a liquid, autoclaving, or steam cleaning. Never
expose cables to strong ultra-violet radiation. Do not sterilize the device or ECG cable with
Ethylene Oxide (EtO) gas.
Summary of Contents for Hillrom H12+
Page 6: ...SERVICE AND SPARE PARTS 2...
Page 8: ...NOTICES 4...
Page 10: ...WARRANTY INFORMATION 6...
Page 18: ...GENERAL CARE 14...
Page 27: ...INTRODUCTION 23 H12 Recorder in Carrying Case...
Page 40: ...CONFIGURING THE RECORDER 36...
Page 46: ...MAINTENANCE 42...
Page 48: ...TROUBLESHOOTING 44...