28
Hygienic preparation
EN
• Clean the humidifier
the relevant operating instructions.
• Clean the bacteria filter
every day according to the
relevant operating instructions.
Change the particle filter whenever there is a
change in patient, after 24 hours' usage at the latest.
5.2 Cleaning
Clean hose system
1. Detach the hose system
2. Pull out one end of the pressure measurement tube
supplied. At the
other end, close the small opening in the adapter
with the second stopper to prevent water enter-
ing.
3. Clean the creased hose so no residues remain with
a small amount of detergent in warm water. Rinse
the inside of the hose thoroughly.
4. Then rinse thoroughly both inside and outside using
clean warm water.
Important!
5. Shake the hose system thoroughly.
6. Hang up the hose system and allow to drip to pre-
vent moisture from getting into SOMNO
smart
2.
The hose system is dried completely with the help of
SOMNO
smart
2:
7. Remove the plugs from the pressure measurement
tube
.
8. If water accidentally gets into the pressure
measurement tube, insert the red drying adapter
supplied into the unit outlet