Tip!
If the granulate manufacturer changes, or the grain size of the granulate changes
significantly, dosing performance determination must be performed again so that the
reserve message is accurate.
Main menu
Settings
Dust extraction (optional)
The dust extraction parameters are set here.
Set flushing interval
Set flushing time
Switch dust extraction on or off
Confirm settings using OK .
Main menu
Settings
Buffer tank
This menu makes it possible to set the dosing performances for filling
the buffer tank. Select the dosing performance so that the required
concentration is present after the buffer tank is filled. For monitoring
purposes, the concentration and pH value must be analysed repeatedly
after the buffer tank has been filled!
The pH value in the buffer tank must be between 6.8 and 7.2.
Off - On:
Activate or deactivate buffer tank filling function
permanently. The filling is regulated automatically via the level control
in the buffer tank.
NOTICE!
When first commissioned, or if the buffer tank is empty, the "Start switch monitoring"
error must be confirmed twice, until the "Start filling" level is reached.
Main Menu
Settings
pH Supervision Buffertank
The parameters for pH monitoring and alarm message are set here.
Alarm low:
Lower alarm value
Alarm high:
Upper alarm value
Delay:
for fine adjustment of the monitoring
pH supervision off / on:
Operating mode pH supervision on / off
Determining dosing performance