CNDAC
5 Operation
Krautkrämer USM 100
209
Sensitivity correction
to com-
pensate for transfer losses in the material to be tested.
A correction is necessary if the test object and the refer-
ence standard have different surface qualities.
You have to determine the adjustment value for the
compensation of transfer losses by way of experiment.
In this connection, the gain is varied accordingly but the
curve shape remains the same.
Adjusting reference lines
You can change the reference lines SL, RL and EL at
any time. The dB values always refer to the reference
line during recording. These values can also be positive,
i.e. positioned above the reference.
and change the Sizing Line as
required.
– With the functions
and
Line and the Evaluation Line in the same way.
Summary of Contents for Krautkramer USM 100
Page 1: ...Krautkr mer USM 100 Operating Manual Id No 160M4395 Rev 9 09 2023...
Page 14: ...14 Rev 9 09 2023 Krautkr mer USM 100 empty page...
Page 21: ...Krautkr mer USM 100 Rev 9 09 2023 21 Introduction 1...
Page 36: ...36 Rev 9 09 2023 Krautkr mer USM 100 empty page...
Page 37: ...Krautkr mer USM 100 Rev 9 09 2023 37 Standard package and accessories 2...
Page 45: ...Krautkr mer USM 100 Rev 9 09 2023 45 Initial start up 3...
Page 66: ...66 Rev 9 09 2023 Krautkr mer USM 100 empty page...
Page 67: ...Krautkr mer USM 100 Rev 9 09 2023 67 Principles of operation 4...
Page 103: ...Krautkr mer USM 100 Rev 9 09 2023 103 Operation 5...
Page 238: ...238 Rev 9 09 2023 Krautkr mer USM 100 empty page...
Page 239: ...Krautkr mer USM 100 Rev 9 09 2023 239 Documentation 6...
Page 257: ...Krautkr mer USM 100 Rev 9 09 2023 257 Maintenance and care 7...
Page 267: ...Krautkr mer USM 100 Rev 9 09 2023 267 Interfaces and Peripherals 8...
Page 287: ...Krautkr mer USM 100 Rev 9 09 2023 287 Appendix 9...
Page 303: ...Krautkr mer USM 100 Rev 9 09 2023 303 Specifications 10...
Page 309: ...Krautkr mer USM 100 Rev 9 09 2023 309 Index 11...
Page 329: ...Krautkr mer USM 100 Rev 9 09 2023 329 empty page...