LYNC UV-H
SECTION 7: MAINTENANCE
•
•
•
7.9 Reassembling the Unit
1. Replace the wiper assembly carefully in the quartz sleeve -wetting the inside of the quartz sleeve
with water will also make the task easier. Turn the wiper assembly counter-clockwise (looking from
the top) as it is being inserted into the quartz sleeve
– this will make the task easier and align the
wiper blades properly. Ensure the bottom of the wiper is correctly seated into bottom manifold.
Figure 7-8
2. Ensure quartz O-ring seal is in place around quartz sleeve and then replace the top manifold by
aligning the shaft adapter into the seal cavity of the top manifold (see Figure 7.8).
3. Tighten the nuts in an alternating pattern. Once top manifold is secure, replace the retaining ring on
the wiper shaft. Rotate the wiper assembly CCW so the top wiper blade faces the back of the unit.
4. Reconnect both top and bottom piping connections. Close the UV Chamber door.
5. Close any taps and slowly open the water supply. Inspect for leaks. Repair any leaks if necessary
before plugging in the unit.
6. Reinstall the motor mounting plate by using the trigger cam to center the plate (center hole to be
concentric with wiper shaft). With the wiper blade opposite the UV sensor facing the back, the
trigger cam should have the dimple in the 10 o
’clock position. Install the wiper motor, then
reconnect all the wires.
7. Reinstall the top and bottom plastic covers.
8. Plug in the unit. Check operation of wiper motor, purge valve and water temperature switch.
9. Make an entry in the service record to establish a cleaning schedule.
7.10 Replacement Parts
Use only genuine parts from Lync when servicing your UV-H disinfection system. Failure to use genuine
replacement parts will void the factory warranty, and any laboratory validation and/or certification for
water safety and system operating performance. Figures 7.9A-B shows a complete list of original factory
parts.
Replacement parts and service are available from Lync representatives.