2.1 Specifications
Measurement Range
0.00 – 15.00 mg/l (PPM)
Accuracy
±10% of reading or ±0.03 mg/l (PPM)
whichever is greater for range of 0 - 15.0 mg/l (PPM)
Resolution
0.01 mg/l (PPM)
Cycle Time
Adjustable; 45 seconds to 10 minutes (600 seconds)
Display
3.5" Color Graphic
Alarms
Two programmable, 120 - 240 VAC 2A, Form C Relay
Analog Output
Powered 4 - 20 mA, 1000Ω drive, isolated
Communications Port
Bi-directional RS-485 with Modbus
Water Pressure
Integral pressure regulator, maximum pressure 3 bar (45 PSI.)
Flow Rate to Waste
Maximum waste flow rate from the drain is 150 ml/min
Operating Temperature
0°C – 55°C (32°F – 131°F)
Wetted Materials
Polyethylene, Borosilicate Glass, Generic Viton™* Alternative (FKM), Polypropylene, 316 Stainless Steel, Acetal, EPDM, Silicone, Modified
PPO, Thermoplastic Elastomer
Sample Temperature Range
0°C – 55°C (32°F – 131°F)
Power Supply
100 to 230VAC, 47 - 63 Hz, 240VA
Insulation Rating
Double Insulated, Pollution Degree 2, Overvoltage Category II
Enclosure
Powder coated steel with windows. IP55 rating
Environmental Conditions
Altitude up to 2000 meters
Up to 95% RH (non-condensing)
IECEx Hazloc Rating
II 2G Ex pxb IIC T4 Gb 0°C ≤ Ta ≤ 55°C
Compressed Air Requirements
Water and oil-free, Particles <5u, ISA Grade Hydrocarbon Free. Full time clean dry air at 5.5 - 7 bar (80 - 101.5 PSI) @ 35 SLPM
(1.2 SCFM) @ 20oC (68oF) Max
Regulatory Compliance And Certifications
IECEx, ATEX, DNVGL-CG-0339, tested to USCG 46 CFR Part 162
subsection 162.060-30
Shipping Weight
27.2 kg (60 lbs.)
Shipping Dimensions
93 cm X 62 cm X 32 cm (36½" X 24½" X 12½")