13
5.2.1.3 Analog Output Menu
Select using UP & DOWN buttons)
SSR-Ex allows the user to determine how an error message will affect
the analog outputs. Settings include driving the analog output either
to 0 mA, 2 mA, or 4 mA or the setting can be OFF to not affect the
output.
Default value is OFF.
SSR-Ex allows the analog output to be scaled to the desired range
using the 4 mA and 20 mA adjustments. Select either 4 mA or 20 mA
and using the UP & DOWN buttons adjust to the desired value.
Default values are 4 mA = 0 ppm and 20 mA = 15 ppm
The 4 and 20 mA outputs can be offset to account for signal loss
between the SSR-Ex and the SCADA system using the 4 mA and 20
mA CALIBRATE options.
ERROR LEVEL
MODE
ALARM 1 TYPE
4 mA
BAUD RATE
ALARM 1 SETPOINT
20 mA
ADDRESS
ALARM 2 TYPE
4 mA CALIBRATE
PARITY
ALARM 2 SETPOINT
20 mA CALIBRATE
STOP BITS
ANALOG OUTPUT
MODBUS
ALARMS
5.2.1.4 Modbus Menu
5.2.1.5 Alarms Menu
The factory default setting is 9600, 8 bit, no parity, 1 stop bit, RTU. All
parameters are adjustable. Make sure the parameters on the Host and
the SSR-Ex are configured identically.
(Select using UP & DOWN buttons)
These settings allow the user to configure the SSR-Ex MODBUS.
Settings include the MODE, BAUD RATE, ADDRESS, PARITY and
BITS.
(Select using UP & DOWN buttons)
SSR-Ex allows the alarms to be configured to be either HI, LO, OFF or
ERROR and the set points can be configured to actuate the relays at
the desired values.
ppm
mg/l
UNITS
(Select using UP & DOWN buttons)
SSR-Ex will display units in either ppm or mg/l. Default value is ppm.
(Select using UP & DOWN buttons)
SSR-Ex averages the selected number of readings and displays the
average on the display MODBUS and on the analog output.
Default value is 2.
(Select using UP & DOWN buttons) Change is in increments of 10.
Default value is 80%.
5.2.1.8 Backlight Menu
5.2.1.7 Average Menu
5.2.1.6 Units Menu
AVERAGE
1 - 5
% RANGE
10% - 100%