CLX-XT (28652)
Page 13
REV 5 (5/2020)
4.4 Installing Reagents
The CLX-XT will require that two reagents be installed prior to operation. These are a
buffer and an indicator. A 12 month supply of reagents can be ordered under catalog no.
09991. For reagent preparation refer to section 10.2
Replacing and Installing the
Reagents
. The reagent life was made under assumption that the CLX-XT would be used
for ballast water monitoring and the system will operate only while the vessel is ballasting
and discharging ballast water.
The Buffer is ready to use and is provided in a 500 ml bottle. The indicator requires some
preparation and is supplied in four 125 ml bottles. Each indicator reagent will last a
maximum of 3 months based on reagent aging when kept in the operating cooling
chamber.