CLX-XT (28652)
Page 24
REV 5 (5/2020)
7.13
4mA Adjustment
If the 4-20 mA setting is turned ON (7.1
Setting the 4-20 mA Output
), the following two
menus will appear. The first menu outputs a constant 4
mA while allowing for a small amount of adjustment.
The adjustment can be made using the
and
buttons.
This adjustment will allow the operator to make the
CLX-XT agree with a PLC or SCADA system. The
adjustment limits are ± 200 counts or about ± 0.2 mA.
This setting will be slightly different on each instrument as each CLX-XT will be factory
set to 4.00mA. Press the
↵
button when adjustments are complete to save this setting and
move on to the 20mA adjustment.
7.14
20mA Adjustment
This menu operates similar to the previous menu. This menu outputs a constant 20 mA
while allowing for a small amount of adjustment. The adjustment can be made using the
and
buttons. The adjustment limits are ± 1000 counts or about ± 1 mA.
This setting will be slightly different on each instrument
as each CLX-XT will be factory set to 20.00mA.
When complete with the 20mA adjustment, press the
↵
button to exit to
AUTO
mode and
save all configuration settings.