Auto restart
Press
ENABLE/DISABLE
to turn the auto restart feature on/off.
This pump offers an auto restart feature. If active on power loss, it restores the pump when
power returns to the operating state it was in when power was lost. For example, if the
pump was running in analog mode prior to power loss, it would return to the same
operating mode and continue to run at a proportional speed to the analog input.
This pump offers an auto restart feature. If active on power loss, it restores the pump when
power returns to the operating state it was in when power was lost. For example, if the
pump was running in manual mode prior to power loss, it would return to the same
operating mode and continue to run at the same speed.
If power is lost in the middle of a dose, when the power is returned it will continue and
finish the interrupted dose.
Any pulses that were in memory before power loss will be remembered. Pulses received
during power loss will be lost.
Do not use auto restart for more than 20 mains power starts per hour.
We recommend remote control where a high number of starts is
required.
The
!
icon is displayed on the home screens to indicate that the auto restart feature is
active.
m-qdos-en-02