C-2
Preparing Sample Vials
Preparing vials
For both reference and modifier samples, you must use only closed vials when
placing samples in the heated vial carrier.
Preparing closed vials
Required materials
•
Chemical-resistant, powder-free gloves
•
Ceramic wafer
•
1.5-mL crimp-top autosampler vials
•
Septum caps
•
Deactivated fused silica capillary (0.32-mm, 0.25 internal diameter, or
both)
•
Vial crimper
•
Syringe
To prepare closed vials:
1. Place approximately 100 µL of sample in a crimp-top autosampler vial.
2. Use the vial crimper to secure a septum cap on the vial.
3. Use the ceramic wafer to cut a 1-cm length of deactivated fused silica
capillary.
4. Use the syringe to puncture the septum cap.
5. Insert the 1-cm length of deactivated fused silica capillary through the
hole in the septum cap, leaving a 0.5-cm length protruding from the vial.
Warning:
To avoid puncture and laceration injuries, take care
when handling the capillary, which is fragile and easily broken.