APGC components
1-7
Disengaging the APGC trolley:
To disengage the APGC trolley:
Requirement:
Before performing this procedure, ensure that you have first
loosened the stabilizers under the trolley.
1. Lift the locking lever.
2. Pull the locking tab outward as far as it will go.
3. Drop the locking lever so that the outermost notch engages with the
locking tab.
4. Carefully slide the trolley away from the locking plate.
5. Pull the locking plate away from the instrument.
1
2
4
3
5
Supporting pillars
(trolley)
Locking plate
Locking lever
Locking tab
Supporting pillars
(instrument)