Maintenance
3
Required Materials
• Slotted head screwdriver
• Powder-free gloves
Procedure
To replace the lamp:
1. Power off the 2996 Detector, disconnect the power cord, and allow the lamp to cool
for at least 15 minutes.
2. Lift up the front panel cover and pull it away from the chassis.
3. Open lamp secure panel.
Figure 3-4 Lamp Power Connector and Mounting Screws
Caution:
To avoid electrical hazards when you perform the following procedure, power
off the 2996 Detector and disconnect the power cord.
Caution:
To avoid the possibility of contacting hot surfaces, wait at least
15 minutes after powering off the detector before you handle the lamp.
TP01466
Alignment
Notch
Mounting
Screws
Lamp Power
Connector
Lamp
Lamp Usage
Indicator