background image

30119-41      Rev. 1.6/02-07

3

MODEL TR28-2

DIGITAL INDICATOR-TOTALIZER-TRANSMITTER

INDEX

I. DESCRIPTION

II. SPECIFICATIONS

III. UNPACKING

IV. INSTALLATION

V. METERS

1.

Remove

2.

Clean Meter Head

3.

Transmitter Mounting Base

4.

Postition of Indicator-Totalizer-Transmitter

5.

Transmitter Wiring

VI. OPERATION AND MAINTENANCE MANUAL

VII. PRINCIPALS OF OPERATION

1.

TR28-2 Functions

2.

Battery

3.

Power to TR28-2

VIII. SENSOR and TR28-2 DIGITAL

INDICATOR-TOTALIZER-TRANSMITTER

1.

TR28-2 Digital Transmitter

2.

TR28-2

3.

Sensor Assembly

4.

Battery Replacement

IX. TROUBLESHOOTING

1

. Troubleshooting Guide

2.

Working Area

3.

TR28-2 Digital Transmitter

4.

Moisture

5.

Transmitter Communication Lines

6.

TR28-2 Circuit

X. INSPECTION

XI. REASSEMBLY

XII. INOPERATIVE INSTRUMENT EVALUATION

1

. Check All Instruments and Transmitters

2.

Check Your Instrument

3.

Transmitter or Communication Line

4.

Full Scale Output

5.

Transmitter Communication Lines

6.

Communications Problems

7.

Separation of Signal and Power

I.

DESCRIPTION:
The Model TR28-2 Digital Indicator Totalizer
Transmitter

 provides flow rate indication, totalization

of flow volume and a current output signal proportional
to the rate of flow when mounted on our meters.  The
TR28-2 is for installation on propeller meters.

II. SPECIFICATIONS:

ACCURACY Rate

±0.25% of reading.

TEMPERATURE

Operation: 32° to 160° F.

RANGE

Storage:

-40° to 160° F.

Consult factory for special construction for other
temperatures.

INPUT SIGNAL

Type:

Voltage pulse.

Voltage Range:

1 to 10V.

Minimum Frequency:

0.125Hz.

Maximum Frequency:

3KHz.

Minimum Pulse Width:

2

μ

s.

OUTPUT SIGNALS

Combination 4-20mA Output and
Isolated Scaled Pulse Output:
Standard 2 wire loop powered.
16 bits resolution.
Operating voltage:

12 to 32 VDC.

Power Switch:  Auto battery override.

Scaled Pulse Output:
Open collector MOS transistor.

Pulse width:

32 milliseconds.

Maximum Rating:

1 to 32 V.

EMI/EMC

Electrostatic Discharge:

8KV

(IEC 1000-4-2 Level 3).

Electrical Fast Transient:

1KV

(IEC 1000-4-4 Level 3).

RF Susceptibility:

150 KHz to 230 MHz @ 10V
(IEC 1000-4-4 Level 3).

POWER SUPPLY

24 VDC (as supplied by our power supply Model

IN36-1, available separately) wired in series.
The Indicator-totalizer-transmitter is primarily
loop powered, but becomes self-powered via a
lithium battery in the event of a power loss.

Battery Type - 3 VDC Lithium,
Replaceable
Operating Life - 2 years (when used
with the display timeout into sleep mode
feature).
Low Battery Indication - 6 months before
expiration.

OPTIONAL

Mounting brackets, with up to 100 feet

EQUIPMENT

of cable for remote installation.

SHIPPING WEIGHT

4 pounds.

ORDERING INFO

Must be specified by the customer and includes
serial number of meter on which unit is to be
mounted, maximum scale range required for 4-20
output, indicator scale and units, totalizer dial
units.  Consult factory for special applications.

Summary of Contents for TR28-2

Page 1: ...MANUAL PARTS LIST FEATURING MODEL CN08 2 DIGITAL INDICATOR TOTALIZER ENCAPSULATED ELECTRONICS SOLID STATE CONSTRUCTION CURRENT OUTPUT SIGNAL PULSE OUTPUT SIGNAL SEALED HOUSING 3255 WEST STETSON AVENU...

Page 2: ...arantees or liabilities arising by law or otherwise Seller shall not incur any other obliga tions or liabilities or be liable to buyer or any customer of buyer for any anticipated or lost profits inci...

Page 3: ...TEMPERATURE Operation 32 to 160 F RANGE Storage 40 to 160 F Consultfactoryforspecialconstructionforother temperatures INPUT SIGNAL Type Voltage pulse Voltage Range 1 to 10V Minimum Frequency 0 125Hz M...

Page 4: ...e wiring connec tions must be made to the transmitter Connect the wire from the transmitter output cable 13 to the mA output terminal on the bottom of the unit The output is polarized Be sure terminal...

Page 5: ...24 VDC loop power to the Model TR28 2 is now on then use troubleshooting pro cedures 2 3 5 6 X XI and XII B If the meter indicator totalizer and remote in strument do not operate check the TR28 2 by...

Page 6: ...CHECK YOUR INSTRUMENT to be sure it is op erating properly by following the instrument service manual Check to be sure your instrument is operating correctly when a simulated signal is put into it WAR...

Page 7: ...sewhere Sometimes it is necessary to lay another line from the transmitter to instrument above ground to help prove the lines are at fault prior to pulling and replacing existing line Communications l...

Page 8: ...ER RESET WIRE IF DESIRED SEE IN16 DATA SHEET WHITE BLACK WHITE BLACK PULSE OUTPUT WIRING WHEN USED WITH AUXILIARY EQUIPMENT PULSE OUTPUT WIRING WHEN USED WITH MODEL IN16 MODEL TR28 2 MODEL TR28 2 Isol...

Page 9: ...size and the power supply Based on a 24V DC power supply and 22 gauge wire we recommend a maximum loop of 3 500 feet not used 4 20 mA Output Left Pin Right Pin Green Red Sensor Wiring Input Left Pin R...

Page 10: ...4 20 mA OUTPUT WIRED TO A MODEL IN48 12 CHART RECORDER SOLD BEFORE 05 93 250 OHM RESISTOR JUMPER RED GREEN AMMETER 4 20 mA NOTE Connections for checking transmitter power supply voltage 24 VDC Connect...

Page 11: ...R FROM RECORDER NOTE TRANSMITTER CABLE SHIELD TO GROUND JUMPER GREEN J2 PLUG 4 20 mA J11 PLUG TRANSMITTER POWER 24 VDC 4 20 mA Output Left Pin Right Pin Green Red TB 2 GREEN RED Isolated Pulse Output...

Page 12: ...8 1 4318 5 1 1551 17 4 4141 1 1 1115 10 10B 1 1551 18 3 1701 4 1 1711 5 No 1 2 3 4 5 6 8 9 10 11 12 13 14 Description Digital Indicator Totalizer Bonnet Assembly O ring Digital Indicator Totalizer Bon...

Page 13: ...30119 41 Rev 1 6 02 07 13 DIGITAL INDICATOR TOTALIZER TRANSMITTER MODEL TR28 2 FOR ELECTRONIC PROPELLER METERS...

Reviews: