background image

Copy for the user / customer

CER

TIFICA

TION

OF

TRAINING

essential for EU user /

customer

The user / customer has been trained in all functions of the sur

gical unit in

accor

dance with curr

ent Instructions for Use. Particular attention was shown to

Safety notes, Disinfecting, Cleaning, Sterilization and Servicing.

W&H

Sur

gical unit

T

ype

SN

wh.com

Name of the user / customer

Clinic, Department

Addr

ess

Signatur

e

Name of the instructor

Addr

ess

Date

Signatur

e

Summary of Contents for OsseoSet 200

Page 1: ...Instructions for use SI 923 SI 915...

Page 2: ...rsons or objects W H Service Sterilizable up to the stated temperature Thermo washer disinfectable Only for USA Caution Federal law restricts this device to sale by or on the order of a dentist physic...

Page 3: ...f motor with cable 14 8 Starting operation foot control 15 20 9 Starting operation General 21 10 Factory settings 22 23 11 Control unit operation 24 27 12 Error messages 28 29 13 Disinfection Cleaning...

Page 4: ...he safety notes on Pages 8 to 11 Intended use Mechanical drive unit with coolant supply for transmission instruments with coupling system according to ISO 3964 DIN 13 940 The equipment is a drive unit...

Page 5: ...for the safety reliability and performance of the OsseoSetTM 200 when there is compliance with the following directions OsseoSetTM 200 must be used in accordance with these Instructions for Use OsseoS...

Page 6: ...ial recycling companies However we recommend that you keep the original packaging Lift out insert with stand Remove foot control and insert Lift out insert with control unit Remove carton with mains c...

Page 7: ...20016 Mains cable REF NB01343700 Europe REF NB03212700 UK IRL REF NB02909300 AUS NZ Motor support REF NB05028500 Stand REF NB04005900 Locking pins REF NB04006800 2 pcs Clips REF NB04019000 5 pcs Drill...

Page 8: ...for surgical straight and contra angle handpieces with respect to maximum speed maximum torque forward and reverse movement Inappropriate use Improper use in addition to incorrect assembly installatio...

Page 9: ...n against the ingress of water The OsseoSetTM 200 Programs 4 and 5 are only designed to be used with the contra angle handpieces WI 75 E KM and WS 75 E KM approved by W H Use of other contra angle han...

Page 10: ...rograms the last values set are saved and re activated on power up System failure A total system failure of the OsseoSetTM 200 does not constitute a critical fault Intermittent operating mode S3 4min...

Page 11: ...gy Fast deceleration of the bur can at times cause the selected torque to be overloaded as a result of the rotational energy stored in the drive system We therefore advise compliance with the individu...

Page 12: ...12 5 Description of front panel Motor connection socket Display OPEN Pump arm P5 Program buttons P1 P2 P3 P4 Pump arm Stand holder...

Page 13: ...13 6 Description of rear panel Connecting socket for foot control Fuse holder with 2 fuses REF 04014700 250 V 1 25A slow blow Power socket Power switch ON OFF Stand holder...

Page 14: ...g the transmission the locking pin supplied can be pushed into one of the two holes see illustration The locking pins can only be used in combination with straight and contra angle handpieces that hav...

Page 15: ...e foot control by 180 Remove the protective cap from the handle hole Locate the guide pin of the handle in the recess in the hole within the foot control Press the handle until it audibly engages into...

Page 16: ...rol removing the pedal for disinfection and cleaning see Page 30 Remove the handle from the foot control Press on one of the two guide pins with the tool provided and lift the pedal slightly Press on...

Page 17: ...l Make sure that both springs are inserted in the appropriate recess Position the pedal at the end of the foot control into the mounting plate Press the two guide pins in front of the foot control at...

Page 18: ...of pedal LIGHT GREY Pump ORANGE Change program DARK GREY Change motor direction BLACK Pedal Motor start Function BLACK VARIABLE or ON OFF Factory setting variable LIGHT GREY Pump ON OFF ORANGE Progra...

Page 19: ...tion Pump ON OFF Only when the motor is at a complete standstill can the pump be switched on or off by operating the LIGHT GREY button of the foot control When pump operation is active the pump symbol...

Page 20: ...ed throughout this procedure Press P3 for approx 4 seconds Keep P3 depressed and simultaneously press PLUS and MINUS Keep P3 depressed and carry out adjustment 01 VARIABLE factory setting Press PLUS b...

Page 21: ...ng operation General Connect the mains cable and foot control Connect the motor cable Note the positioning Insert the stand Note the positioning Fit the motor support Fit the irrigation tubing set Clo...

Page 22: ...ettings Switch off the control unit Keep P1 depressed and simultaneously switch on the control unit Keep P1 depressed as long as on the display the adjustment DE FAU appears Factory setting always sta...

Page 23: ...on forward forward forward Pump on on on Factory setting Ncm max max max Adjustable range Ncm P4 forward P4 reverse P5 forward P5 reverse Transmission ratio 20 1 20 1 20 1 20 1 Factory setting rpm 30...

Page 24: ...an audible signal can be heard and the Program button is illuminated On the display the selected Program appears with the adjusted range in rpm e g for P1 rpm Ncm rpm Display adjustments P1 P3 speed...

Page 25: ...l unit operation Changing the speed P1 P3 Keeping PLUS MINUS depressed activates the repeat function and the values are continuously increased decreased Press Program button P1 P3 Increase speed Decre...

Page 26: ...tion and the values are continuously increased decreased When changing from 5 to 50 Ncm and 50 to 5 Ncm a longer confirmation signal is heard When the adjustable torque in the the forward reverse driv...

Page 27: ...ously increased decreased Keep P2 depressed for approx 4 seconds the adjusted coolant flow rate appears Keep P2 depressed and press PLUS button to increase flow rate Keep P2 depressed and press MINUS...

Page 28: ...start do not initializing actuate foot control when switching on 09 Foot control error Switch off the equipment check connection of the foot control then re start 19 Run limiting control Switch off t...

Page 29: ...tching off the OsseoSetTM 200 and then switching it on again the equipment must be checked by a service and repair organization authorized by W H see Page 45 47 If a total failure of the equipment occ...

Page 30: ...al disinfection and cleaning The front panel of the control unit is sealed and may be wiped clean Remove handle from the foot control Remove pedal from the foot control see Page 16 If heavily soiled C...

Page 31: ...is recommended Use certified surface disinfectants e g DGHM tested For USA Use EPA registered surface disinfectants Note the action times of the disinfectants used Manual cleaning Rinse under deminera...

Page 32: ...rotective cover Steam vacuum sterilization according to EN 13060 using a sterilization holding time of a minimum of 3 minutes at 134 3 C 273 2 5 4 F or Gravity sterilization Remove the motor protectiv...

Page 33: ...le For spare parts and other accessories not listed here please contact your local W H representative or service station NB02610500 Cannula for internal bur cooling NB04006800 Locking pin NB10207554 C...

Page 34: ...rtaken by a qualified organization and needs to include the following procedures visual inspection for outside damage measurement of enclosure leakage current measurement of patient leakage current vi...

Page 35: ...always return all the equipment due to the fact that with motor malfunctions an inspection of the electronic controls is also necessary Returns Refer all questions to your authorized dental dealer or...

Page 36: ...S3 4min 10min Dimensions in mm W x D x H 252 x 254 x 90 252 x 254 x 90 Weight in kg 3 30 3 30 The foot control is water tight according to IPX8 1m depth of immersion 1 hour water tight according to I...

Page 37: ...of the OsseoSetTM 200 equipment The equipment contains many valuable materials Therefore return your equipment for material recycling via the relevant public collection system Disposal of the foot con...

Page 38: ...uring defects within a warranty period of one year from the date of purchase We accept no responsibility for damage caused by incorrect handling or by repairs carried out by third parties not authoriz...

Page 39: ...all functions of the surgical unit in accordance with current Instructions for Use Particular attention was shown to Safety notes Disinfecting Cleaning Sterilization and Servicing W H Surgical unit T...

Page 40: ...ned in all functions of the surgical unit in accordance with current Instructions for Use Particular attention was shown to Safety notes Disinfecting Cleaning Sterilization and Servicing W H Surgical...

Page 41: ...mail office ch wh com W H Dentalmechanik GmbH Ignaz Glaser Stra e 60 A 5111 B rmoos Tel 43 0 6274 6236 239 Fax 43 0 6274 6236 890 E mail service dept wh com CH A D W H FRANCE S A R L 4 rue Ettore Buga...

Page 42: ...6 S 187 66 T by Tel 46 0 8 445 88 30 Fax 46 0 8 445 88 33 E mail office whnordic se I CZ WEHADENT IB RICA S L C Ciudad de Melilla 3 Bajo E 46017 Valencia Tel 34 96 353 2020 Fax 34 96 353 2579 E mail...

Page 43: ...969 946 00 Fax 61 2 969 947 00 E mail adec adec com au SHINWON DENTAL CO LTD 158 Namdaemun ro 5 ka Joong ku Seoul Phone 82 2 774 68 51 Fax 82 2 310 96 64 E mail phkwon shinwon dental com W H Impex 74...

Page 44: ...Box 5190 SE 402 26 G teborg SWEDEN Nobel Biocare USA LLC CA USA Distributed in North and South America by Manufacturer ISO 13485 Annex II 93 42 EWG EC 0297 W H Dentalwerk B rmoos GmbH Austria A 5111 B...

Reviews: