82
Description
The Helix test device is used to validate the sterilizer performance for
hollow items.
It consists of a 1,500 mm long tube that is open on one side and closed
with a capsule on the other side.
A chemical indicator strip is placed inside of the capsule.
How to carry out the test
The test must be performed in an empty chamber (EN13060) without load but with the standard chamber accessories (chamber rack
and trays) mounted.
Place an indicator strip inside the capsule according to the instructions of the test manufacturer.
Close the capsule.
Once the cycle is finished, remove the test device from the chamber.
Remove the indicator strip from the capsule and check the change in colour:
TEST PASSED
The indicator strip has turned dark.
The air was completely removed from the capsule.
TEST FAILED
Part of the chemical indicator strip has not turned dark; showing
that the air removal from the capsule was not complete.
If the test fails repeatedly call technical service.
Follow local/national guidelines on the frequency of testing.
ANNEX 5. Helix test
Start the cycle: from the CYCLE SELECTION screen (HOME screen) select HELIX B&D TEST (see “Running a sterilization cycle”).