72
The sterilizer is fitted with a safety thermostat to prevent overheating.
If the safety thermostat operates because of too high temperatures, the Error
240 or a timeout error is displayed.
The thermostat must be reset manually. Proceed as follows:
- Switch the sterilizer OFF and remove the mains cable.
- Wait for the sterilizer to cool down.
- Remove the dust filter.
- Slide your hand underneath the sterilizer where the dust filter was located and
push on the reset button of the thermostat switch (see pictures to the left).
-A click sound indicates that the thermostat switch has been reset.
- Place the sterilizer back into its original position.
- Insert the dust filter back into its original position.
- Connect the mains cable and switch the sterilizer ON.
- Wait for the sterilizer to finish the error reset phase and follow the instructions
on the display.
If the thermostat operates repeatedly, call technical service.
Resetting the safety thermostat
For better access to the reset button, move the sterilizer closer to
the shelf edge.