Hygiene and maintenance
Sterilization
Evidence of the medical device’s basic suitability for effective sterilization was provided by an independent test laboratory
using the LISA 517 B17L* steam sterilizer (W&H Sterilization S.r.l., Brusaporto (BG)) and the CertoClav MultiControl MC2-
S09S273** steam sterilizer (CertoClav GmbH, Traun).
“Dynamic-air-removal cycle”: 132°C (270°F) – 4 minutes*/**
“Gravity-displacement cycle”: 121°C (250°F) – 30 minutes**
Minimum drying times:
“Dynamic-air-removal cycle”: 132°C (270°F) – 30 minutes**
“Gravity-displacement cycle”: 121°C (250°F) – 30 minutes**
* EN 13060, EN 285, ISO 17665
** ANSI/AAMI ST55, ANSI/AAMI ST79