> If the described problem cannot be resolved, the unit will need to be inspected by an authorized W&H service partner.
> In case of a total system failure, switch the control unit off and on again.
Icon
Description of error
Solution
WARNUNG CONNECTION
> Press the ioDent® Wi-Fi dongle symbol
> Attempt to establish a connection with the ioDent® platform again.
WARNING DATA RECEPTION
> Restart the data transfer on the ioDent® platform.
WARNING TIME SYNCHRONISATION
> Restart the gateway
> Insert the ioDent® Wi-Fi dongle again
WARNING SYSTEM MONITORING
> Release the pedal (grey) on the foot control and press it again.
> If the error occurs again, restart the device.
WARNING IMPLANT DOCUMENTATION
> Maximum number of implants (8) for the active documentation has
been reached.
WARNING DOCUMENTATION ACTIVE
> Finish the current documentation on the device before starting
a new one.
WARNING SOFTWARE UPDATE FAILED
> Check the update files and copy the data to the USB stick again.
> Insert the USB stick again. Restart the update.
Error messages